回声跟踪法通过测量肱动脉反应性提高了血流介导的血管舒张功能评估的可靠性。

Echo-Tracking Method Improves the Reliability in Flow-Mediated Dilation Assessment by Measuring Brachial Artery Reactivity.

作者信息

Du Li-juan, He Wen, Wang Jian-hua, Gao Yu, Liang Xiao-lu, Zhang Jiang-xia

机构信息

*Department of Ultrasound, Beijing Tiantan Hospital, Capital Medical University; and †Department of Ultrasound, The Military General Hospital of Beijing, Beijing, China.

出版信息

Ultrasound Q. 2015 Sep;31(3):180-4. doi: 10.1097/RUQ.0000000000000129.

Abstract

The purpose of the study was to compare the intraobserver and interobserver reliability and agreement for measurement of flow-mediated dilation (FMD) between B-mode method and echo-tracking (ET) method. Twenty healthy volunteers (mean age, 31.6 ± 9.2 years) underwent ultrasound examination by both B-mode and ET methods. Baseline brachial artery diameter, post-cuff release diameter, and FMD percent were assessed by each sonologist on 2 consecutive days. Reliability was assessed by intraclass correlation coefficients, and Bland-Altman plots were used to visually compare measurement bias and agreement by the 2 ultrasound methods. A total of 40 pairs of data were available for analysis. Excellent intraobserver and interobserver reliability values were found for all variables assessed by the 2 methods. The intraclass correlation coefficient values were higher for ET in both intraobserver and interobserver evaluations, but only for interobserver evaluations for post-cuff release diameter and FMD was there no overlap in the 95% confidence interval. The Bland-Altman plots showed that in 95% of the measurements, the percentage difference between ET and B-mode ultrasound techniques was within 18.1%, 19.4%, and 17.3% for baseline brachial artery diameter, post-cuff release diameter, and FMD percent, respectively. The results suggest that ET and B-mode methods are reproducible in assessing the FMD. The ET method improves the reliability of FMD assessment, but we cannot determine which measurement is better for FMD.

摘要

本研究的目的是比较B超方法和回声跟踪(ET)方法在测量血流介导的血管舒张功能(FMD)时观察者内和观察者间的可靠性及一致性。20名健康志愿者(平均年龄31.6±9.2岁)接受了B超和ET两种方法的超声检查。每位超声科医生在连续两天评估基线肱动脉直径、袖带松开后的直径以及FMD百分比。通过组内相关系数评估可靠性,并使用Bland-Altman图直观比较两种超声方法的测量偏差和一致性。共有40对数据可供分析。两种方法评估的所有变量均发现了出色的观察者内和观察者间可靠性值。在观察者内和观察者间评估中,ET的组内相关系数值均较高,但仅在观察者间评估中,袖带松开后的直径和FMD在95%置信区间内没有重叠。Bland-Altman图显示,在95%的测量中,ET和B超技术在基线肱动脉直径、袖带松开后的直径以及FMD百分比方面的差异百分比分别在18.1%、19.4%和17.3%以内。结果表明,ET和B超方法在评估FMD方面具有可重复性。ET方法提高了FMD评估的可靠性,但我们无法确定哪种测量方法对FMD更好。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验