This work describes a commercial treatment planning system, its technical features, and its capabilities for creating (60)Co intensity modulated radiation therapy (IMRT) treatment plans for a magnetic resonance image guidance radiation therapy (MR-IGRT) system.
The ViewRay treatment planning system (Oakwood Village, OH) was used to create (60)Co IMRT treatment plans for 33 cancer patients with disease in the abdominal, pelvic, thorax, and head and neck regions using physician-specified patient-specific target coverage and organ at risk (OAR) objectives. Backup plans using a third-party linear accelerator (linac)-based planning system were also created. Plans were evaluated by attending physicians and approved for treatment. The (60)Co and linac plans were compared by evaluating conformity numbers (CN) with 100% and 95% of prescription reference doses and heterogeneity indices (HI) for planning target volumes (PTVs) and maximum, mean, and dose-volume histogram (DVH) values for OARs.
All (60)Co IMRT plans achieved PTV coverage and OAR sparing that were similar to linac plans. PTV conformity for (60)Co was within <1% and 3% of linac plans for 100% and 95% prescription reference isodoses, respectively, and heterogeneity was on average 4% greater. Comparisons of OAR mean dose showed generally better sparing with linac plans in the low-dose range <20 Gy, but comparable sparing for organs with mean doses >20 Gy. The mean doses for all (60)Co plan OARs were within clinical tolerances.
A commercial (60)Co MR-IGRT device can produce highly conformal IMRT treatment plans similar in quality to linac IMRT for a variety of disease sites. Additional work is in progress to evaluate the clinical benefit of other novel features of this MR-IGRT system.
本研究介绍了一种商业化的治疗计划系统,描述了其技术特点,并展示了其为磁共振影像引导放射治疗(MR-IGRT)系统创建钴-60 调强放射治疗(IMRT)计划的能力。
使用 ViewRay 治疗计划系统(俄亥俄州奥克伍德村)为 33 例患有腹部、盆腔、胸部和头颈部疾病的癌症患者创建钴-60 IMRT 治疗计划,这些计划是根据医生指定的患者特定靶区覆盖和危及器官(OAR)目标制定的。还创建了使用第三方直线加速器(linac)-based 计划系统的备用计划。由主治医生对计划进行评估并批准用于治疗。通过评估 100%和 95%处方参考剂量的靶区适形度指数(CN)以及计划靶区(PTV)和 OAR 的不均匀性指数(HI)、最大剂量、平均剂量和剂量-体积直方图(DVH)值来比较钴-60 和直线加速器计划。
所有钴-60 IMRT 计划都实现了与直线加速器计划相似的 PTV 覆盖和 OAR 保护。对于 100%和 95%处方参考等剂量线,钴-60 的 PTV 适形度分别比直线加速器计划低<1%和 3%,不均匀性平均高 4%。OAR 平均剂量的比较显示,在低剂量范围<20Gy 时,直线加速器计划通常能更好地保护 OAR,但对于平均剂量>20Gy 的器官,保护效果相当。所有钴-60 计划 OAR 的平均剂量均在临床耐受范围内。
一种商业化的钴-60 MR-IGRT 设备可以为多种疾病部位生成高质量的与直线加速器 IMRT 相似的高度适形 IMRT 治疗计划。目前正在进行更多的工作来评估该 MR-IGRT 系统的其他新特性的临床获益。