Rosen B S, Soares C G, Hammer C G, Kunugi K A, DeWerd L A
Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin 53705.
National Institute of Standards and Technology, Gaithersburg, Maryland 20899.
Med Phys. 2015 Jul;42(7):4055-68. doi: 10.1118/1.4922134.
To evaluate a prototype densitometer traceable to primary optical standards and compare its performance to an EPSON Expression(®) 10000XL flatbed scanner (the Epson) for quantitative radiochromic film (RCF) dosimetry.
A prototype traceable laser densitometry system (LDS) was developed to mitigate common film scanning artifacts, such as positional scan dependence and high noise in low-dose regions, by performing point-based measurements of RCF suspended in free-space using coherent light. The LDS and the Epson optical absorbance scales were calibrated up to 3 AU, using reference materials calibrated at a primary standards laboratory and a scanner calibration factor (SCF). Calibrated optical density (OD) was determined for 96 Gafchromic(®) EBT3 film segments before and after irradiation to one of 16 dose levels between 0 and 10 Gy, exposed to (60)Co in a polymethyl-methacrylate (PMMA) phantom. The sensitivity was determined at each dose level and at two rotationally orthogonal readout orientations to obtain the sensitometric response of each RCF dosimetry system. LDS rotational scanning dependence was measured at nine angles between 0°and 180°, due to the expected interference between coherent light and polarizing EBT3 material. The response curves were fit to the analytic functions predicted by two physical response models: the two-parameter single-hit model and the four-parameter percolation model.
The LDS and the Epson absorbance measurements were linear to primary optical standards to within 0.2% and 0.3% up to 2 and 1 AU, respectively. At higher densities, the LDS had an over-response (2.5% at 3 AU) and the Epson an under-response (3.1% and 9.8% at 2 and 3 AU, respectively). The LDS and the Epson SCF over the applicable range were 0.968% ± 0.2% and 1.561% ± 0.3%, respectively. The positional scan dependence was evaluated on each digitizer and shown to be mitigated on the LDS, as compared to the Epson. Maximum EBT3 rotational dependence was found to have a strong dependence on dose (0.1% and 34% at 30 mGy and 5 Gy, respectively). The preferred EBT3 polymerization axis angle was constant within experimental uncertainties. In its most sensitive orientation, the LDS-measured EBT3 sensitivity was 7.13 × 10(-4) ± 9.2 × 10(-6) AU/mGy, which represented a 4.5 fold increase over the Epson of 1.58 × 10(-4) ± 9.8 × 10(-6) AU/mGy. To first order approximations, EBT3 response was linear up to 500 mGy to within 0.80% and to within 7.5% for the most sensitive LDS and the Epson orientations, respectively. The corresponding single-hit and percolation model relative residual norms were 0.082 and 0.074 for LDS as compared to 0.29 and 0.18 for the Epson, which represented a significant increase in LDS-measured agreement with the simple physical model. Less sensitive LDS and the Epson orientations showed a marked decrease in the physical model agreement, which suggested that suboptimal readout device characteristics may be the origin of the complex sensitometric functional forms currently required for accurate RCF dosimetry.
The prototype densitometer was shown to be superior to a conventional scanner for quantitative RCF dosimetry based on physical models of film response. The Epson was shown to be a reliable tool for routine RCF dosimetry in a clinical setting, yet calibration to primary optical standards did not mitigate the necessity for complex, empirical functional form fitting.
评估一款可溯源至初级光学标准的原型密度计,并将其性能与爱普生Expression(®) 10000XL平板扫描仪(爱普生)用于定量放射变色薄膜(RCF)剂量测定的性能进行比较。
开发了一种可溯源的激光密度测量系统(LDS),通过使用相干光对自由空间中悬浮的RCF进行基于点的测量,以减轻常见的薄膜扫描伪影,如位置扫描依赖性和低剂量区域的高噪声。使用在初级标准实验室校准的参考材料和扫描仪校准因子(SCF),将LDS和爱普生的吸光度标度校准至3 AU。在聚甲基丙烯酸甲酯(PMMA)体模中,将96个Gafchromic(®) EBT3薄膜段照射至0至10 Gy之间的16个剂量水平之一前后,测定其校准光密度(OD)。在每个剂量水平以及两个旋转正交读出方向上测定灵敏度,以获得每个RCF剂量测定系统的感光响应。由于相干光与偏振EBT3材料之间预期的干涉,在0°至180°之间的九个角度测量LDS的旋转扫描依赖性。将响应曲线拟合到由两个物理响应模型预测的解析函数:双参数单击中模型和四参数渗流模型。
LDS和爱普生的吸光度测量值与初级光学标准的线性度分别在2 AU和1 AU以内,误差分别在0.2%和0.3%以内。在更高密度下,LDS有过响应(3 AU时为2.5%),爱普生有欠响应(2 AU和3 AU时分别为3.1%和9.8%)。在适用范围内,LDS和爱普生的SCF分别为0.968%±0.2%和1.561%±0.3%。在每个数字化仪上评估了位置扫描依赖性,结果表明与爱普生相比,LDS上的这种依赖性得到了减轻。发现EBT3的最大旋转依赖性对剂量有很强的依赖性(30 mGy和5 Gy时分别为0.1%和34%)。在实验不确定度范围内,EBT3的首选聚合轴角度是恒定的。在其最敏感方向上,LDS测量的EBT3灵敏度为7.13×10(-4)±9.2×10(-6) AU/mGy,比爱普生的1.58×10(-4)±9.8×10(-6) AU/mGy提高了4.5倍。一阶近似下,对于最敏感的LDS和爱普生方向,EBT3响应在500 mGy以内分别在0.80%和7.5%以内呈线性。与爱普生的0.29和0.18相比,LDS的相应单击中模型和渗流模型相对残差范数分别为0.082和0.074,这表明LDS测量值与简单物理模型的一致性有显著提高。LDS和爱普生不太敏感的方向显示出与物理模型的一致性明显下降,这表明次优的读出设备特性可能是目前准确RCF剂量测定所需复杂感光功能形式的根源。
基于薄膜响应的物理模型,该原型密度计在定量RCF剂量测定方面优于传统扫描仪。爱普生被证明是临床环境中常规RCF剂量测定的可靠工具,但校准至初级光学标准并不能减轻对复杂经验函数形式拟合的必要性。