Morrison Hali, Menon Geetha, Sloboda Ron S
Department of Medical Physics, Cross Cancer Institute, Edmonton, Alberta T6G 1Z2, Canada and Department of Oncology, University of Alberta, Edmonton, Alberta T6G 2R3, Canada.
Med Phys. 2014 Jul;41(7):072101. doi: 10.1118/1.4881146.
Radiochromic film dosimetry is typically performed for high energy photons and moderate doses characterizing external beam radiotherapy (XRT). The purpose of this study was to investigate the accuracy of previously established film calibration procedures used in XRT when applied to low-energy, seed-based brachytherapy at higher doses, and to determine necessary modifications to achieve similar accuracy in absolute dose measurements.
Gafchromic EBT3 film was used to measure radiation doses upwards of 35 Gy from 75 kVp, 200 kVp, 6 MV, and (∼28 keV) I-125 photon sources. For the latter irradiations a custom phantom was built to hold a single I-125 seed. Film pieces were scanned with an Epson 10000XL flatbed scanner and the resulting 48-bit RGB TIFF images were analyzed using both FilmQA Pro software andMATLAB. Calibration curves relating dose and optical density via a rational functional form for all three color channels at each irradiation energy were determined with and without the inclusion of uncertainties in the measured optical densities and dose values. The accuracy of calibration curve variations obtained using piecewise fitting, a reduced film measurement area for I-125 irradiation, and a reduced number of dose levels was also investigated. The energy dependence of the film lot used was also analyzed by calculating normalized optical density values.
Slight differences were found in the resulting calibration curves for the various fitting methods used. The accuracy of the calibration curves was found to improve at low doses and worsen at high doses when including uncertainties in optical densities and doses, which may better represent the variability that could be seen in film optical density measurements. When exposing the films to doses > 8 Gy, two-segment piecewise fitting was found to be necessary to achieve similar accuracies in absolute dose measurements as when using smaller dose ranges. When reducing the film measurement area for the I-125 irradiations, the accuracy of the calibration curve was degraded due to the presence of localized film heterogeneities. No degradation in the calibration curves was found when reducing the number of calibration points down to only 4, but with piecewise fitting, 6 calibration points as well as a blank film are required. Variations due to photon energy in film optical density of up to 3% were found above doses of 2 Gy.
A modified procedure for performing EBT3 film calibration was established for use with low-energy brachytherapy seeds and high dose exposures. The energy dependence between 6 MV and I-125 photons is significant such that film calibrations should be done with an appropriately low-energy source when performing low-energy brachytherapy dose measurements. Two-segment piecewise fitting with the inclusion of errors in measured optical density as well as dose was found to result in the most accurate calibration curves. Above doses of 1 Gy, absolute dose measurements can be made with an accuracy of 1.6% for 6 MV beams and 5.7% for I-125 seed exposures if using the I-125 source for calibration, or 2.3% if using the 75 kVp photon beam for calibration.
放射变色薄膜剂量测定法通常用于高能光子和表征外照射放疗(XRT)的中等剂量。本研究的目的是调查先前用于XRT的薄膜校准程序应用于基于种子的低能近距离放疗且剂量较高时的准确性,并确定为在绝对剂量测量中实现类似准确性而需要进行的必要修改。
使用Gafchromic EBT3薄膜测量来自75 kVp、200 kVp、6 MV和(约28 keV)I-125光子源的35 Gy以上的辐射剂量。对于后一种照射,构建了一个定制体模来容纳单个I-125种子。用爱普生10000XL平板扫描仪扫描薄膜片,使用FilmQA Pro软件和MATLAB分析所得的48位RGB TIFF图像。通过合理函数形式确定每个照射能量下所有三个颜色通道的剂量与光密度之间的校准曲线,包括和不包括测量光密度和剂量值中的不确定性。还研究了使用分段拟合、减少I-125照射的薄膜测量面积以及减少剂量水平数量所获得的校准曲线变化的准确性。通过计算归一化光密度值也分析了所用薄膜批次的能量依赖性。
对于所使用的各种拟合方法,在所得校准曲线中发现了细微差异。当在光密度和剂量中纳入不确定性时,校准曲线的准确性在低剂量时提高而在高剂量时变差,这可能更好地代表了薄膜光密度测量中可能出现的变异性。当将薄膜暴露于大于8 Gy的剂量时,发现需要两段分段拟合才能在绝对剂量测量中获得与使用较小剂量范围时相似的准确性。当减少I-125照射的薄膜测量面积时,由于存在局部薄膜不均匀性,校准曲线的准确性降低。当将校准点数量减少到仅4个时,在校准曲线中未发现退化,但对于分段拟合,需要6个校准点以及一张空白薄膜。在剂量高于2 Gy时,发现薄膜光密度中由于光子能量导致的变化高达3%。
建立了一种修改后的程序用于EBT3薄膜校准,以用于低能近距离放疗种子和高剂量照射。6 MV和I-125光子之间的能量依赖性很显著,因此在进行低能近距离放疗剂量测量时,应使用适当的低能源进行薄膜校准。发现包含测量光密度以及剂量误差的两段分段拟合会产生最准确的校准曲线。如果使用I-125源进行校准,对于6 MV束,在剂量高于1 Gy时,绝对剂量测量的准确度可达1.6%,对于I-125种子照射可达5.7%;如果使用75 kVp光子束进行校准,则准确度为2.3%。