Kinkel Stefan, Nadorf Jan, Thomsen Marc N, Heisel Christian, Jahnke Alexander, Kretzer Jan P, Jakubowitz Eike
Laboratory of Biomechanics and Implant Research, Department of Orthopaedics, Traumatology and Paraplegiology, University Hospital Heidelberg, Schlierbacher Landstrasse 200a, 69118, Heidelberg, Germany.
Int Orthop. 2015 Sep;39(9):1819-25. doi: 10.1007/s00264-015-2869-5. Epub 2015 Jul 9.
Femoral defects often make it difficult to achieve sufficient fixation of the stem during revision surgery. No clinical studies comparing modular stems are available and differentiated recommendations are rare. The aim of this study was to compare the fixation of different revision stems in an experimental and standardised manner.
Segmental AAOS type I and III defects were reproduced in four femur pairs and two modular stems of different shape (cylindrical and conical) were implanted. Interfacial stem-bone movements were measured under axial torque application to analyze the stem fixation depending on defect extension.
Both stems showed adequate fixation in AAOS type I defect. The defect extension significantly reduced the fixation of both implants. The fixation pattern changed significantly for the cylindrical-shaped stem but was maintained for the conical-shaped stem.
Shape as well as the extension of femoral defect have an impact on primary fixation. A type I defect seems to be bridgeable for both stems albeit in a different way. In contrast, stem-demanding activities during the healing phase have to be avoided for the conical-shaped stem in a type III defect, whereas the cylindrical-shaped stem has already ceased to be sufficiently stable in this case.
股骨缺损常常使得翻修手术中股骨柄难以实现充分固定。目前尚无比较模块化股骨柄的临床研究,且缺乏差异化的建议。本研究的目的是以实验性和标准化的方式比较不同翻修股骨柄的固定情况。
在四对股骨上重现节段性美国矫形外科医师学会(AAOS)I型和III型缺损,并植入两种不同形状(圆柱形和圆锥形)的模块化股骨柄。在施加轴向扭矩的情况下测量股骨柄与骨之间的界面运动,以分析取决于缺损范围的股骨柄固定情况。
两种股骨柄在AAOS I型缺损中均显示出足够的固定。缺损范围显著降低了两种植入物的固定效果。圆柱形股骨柄的固定模式发生了显著变化,而圆锥形股骨柄的固定模式则保持不变。
股骨缺损的形状和范围对初次固定均有影响。对于两种股骨柄而言,I型缺损似乎都可以跨越,尽管方式不同。相比之下,III型缺损中,愈合阶段需要对股骨柄施加较大应力的活动必须避免用于圆锥形股骨柄,而在这种情况下圆柱形股骨柄已经不再足够稳定。