Bailey Timothy, Bode Bruce W, Christiansen Mark P, Klaff Leslie J, Alva Shridhara
1 AMCR Institute , Escondido, California.
2 Atlanta Diabetes Associates , Atlanta, Georgia .
Diabetes Technol Ther. 2015 Nov;17(11):787-94. doi: 10.1089/dia.2014.0378. Epub 2015 Jul 14.
The purpose of the study was to evaluate the performance and usability of the FreeStyle(®) Libre™ Flash glucose monitoring system (Abbott Diabetes Care, Alameda, CA) for interstitial glucose results compared with capillary blood glucose results.
Seventy-two study participants with type 1 or type 2 diabetes were enrolled by four U.S. clinical sites. A sensor was inserted on the back of each upper arm for up to 14 days. Three factory-only calibrated sensor lots were used in the study. Sensor glucose measurements were compared with capillary blood glucose (BG) results (approximately eight per day) obtained using the BG meter built into the reader (BG reference) and with the YSI analyzer (Yellow Springs Instrument, Yellow Springs, OH) reference tests at three clinic visits (32 samples per visit). Sensor readings were masked to the participants.
The accuracy of the results was demonstrated against capillary BG reference values, with 86.7% of sensor results within Consensus Error Grid Zone A. The percentage of readings within Consensus Error Grid Zone A on Days 2, 7, and 14 was 88.4%, 89.2%, and 85.2%, respectively. The overall mean absolute relative difference was 11.4%. The mean lag time between sensor and YSI reference values was 4.5±4.8 min. Sensor accuracy was not affected by factors such as body mass index, age, type of diabetes, clinical site, insulin administration, or hemoglobin A1c.
Interstitial glucose measurements with the FreeStyle Libre system were found to be accurate compared with capillary BG reference values, with accuracy remaining stable over 14 days of wear and unaffected by patient characteristics.
本研究的目的是评估FreeStyle(®) Libre™ 实时葡萄糖监测系统(雅培糖尿病护理公司,加利福尼亚州阿拉米达)测量组织间液葡萄糖结果与毛细血管血糖结果相比的性能和可用性。
美国四个临床地点招募了72名1型或2型糖尿病研究参与者。在每个上臂背部插入一个传感器,最长使用14天。研究中使用了三个仅经过工厂校准的传感器批次。将传感器测得的葡萄糖值与使用血糖仪内置的血糖仪(血糖参考值)获得的毛细血管血糖(BG)结果(每天约8次)以及在三次临床就诊时(每次就诊32个样本)使用YSI分析仪(俄亥俄州黄泉市黄泉仪器公司)进行的参考测试结果进行比较。向参与者隐瞒传感器读数。
与毛细血管BG参考值相比,结果的准确性得到了验证,86.7%的传感器结果在共识误差网格A区范围内。第2天、第7天和第14天在共识误差网格A区内读数的百分比分别为88.4%、89.2%和85.2%。总体平均绝对相对差异为11.4%。传感器与YSI参考值之间的平均滞后时间为4.5±4.8分钟。传感器准确性不受体重指数、年龄、糖尿病类型、临床地点、胰岛素给药或糖化血红蛋白等因素影响。
与毛细血管BG参考值相比,发现FreeStyle Libre系统测量组织间液葡萄糖的结果准确,在佩戴14天期间准确性保持稳定,且不受患者特征影响。