Ventilator disconnexion alarm failures. The role of ventilator and breathing system accessories.

作者信息

Pryn S J, Crosse M M

机构信息

Shackelton Department of Anaesthetics, Southampton General Hospital.

出版信息

Anaesthesia. 1989 Dec;44(12):978-81. doi: 10.1111/j.1365-2044.1989.tb09201.x.

Abstract

Pressure-sensitive ventilator disconnexion alarms do not always alarm during disconnexion of a discharging compliance ventilator such as Manley Blease, unless accurately adjusted. High flows during disconnexion result in significant pressure generation caused by outflow resistance of catheter mounts, heat and moisture exchangers, capnometer cuvettes, and angled connectors; this may lead to alarm failure because of incorrectly adjusted pressure alarm limits. The exact position of the disconnexion is critical and if the alarm's pressure sensor is placed in either the inspiratory or expiratory limb of the ventilator it makes no difference to its correct function. Nine different heat and moisture exchanges were compared in the same breathing system. Those with 15-mm male connectors generate the highest pressures on disconnexion (1.1 kPa). It is suggested that the low pressure alarm limit is set only marginally below the peak inspiratory pressure, and that it is readjusted for every patient and after every change in ventilation. Most importantly, the alarm should be shown to be functional by a trial disconnexion at the tracheal tube.

摘要

压力敏感型呼吸机断开连接警报在诸如曼利·布利斯(Manley Blease)等排出顺应性呼吸机断开连接期间并非总能发出警报,除非进行精确调整。断开连接期间的高流量会因导管接头、热湿交换器、二氧化碳分析仪比色杯和弯角连接器的流出阻力而产生显著压力;这可能会因压力警报限值调整不当而导致警报失败。断开连接的确切位置至关重要,并且如果警报的压力传感器放置在呼吸机的吸气或呼气支路上,这对其正确功能没有影响。在同一呼吸回路中比较了九种不同的热湿交换器。那些带有15毫米公接头的在断开连接时产生的压力最高(1.1千帕)。建议将低压警报限值仅设置在略低于吸气峰值压力的水平,并且针对每个患者以及每次通气变化后都重新进行调整。最重要的是,应通过在气管导管处进行一次断开连接试验来证明警报功能正常。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验