Kun S S, Nakamura C T, Ripka J F, Davidson Ward S L, Keens T G
Division of Pediatric Pulmonology, Childrens Hospital Los Angeles, Keck School of Medicine of the University of Southern California, Los Angeles, CA 90027, USA.
Chest. 2001 Feb;119(2):562-4. doi: 10.1378/chest.119.2.562.
Positive-pressure ventilators are equipped with low-inspiratory-pressure alarms to protect patients from hypoventilation. Small uncuffed tracheostomy tubes have a high resistance, and may not trigger these alarms during decannulation.
To determine whether ventilator low-inspiratory-pressure alarms are effective in detecting tracheostomy decannulation.
We connected tracheostomy tubes of varying inner diameters (3.0 to 6.0 mm) to a home ventilator and simulated decannulation using low (tidal volume [VT], 600 mL; peak inspiratory pressure [PIP], 25 cm H(2)O), medium (VT, 800 mL; PIP, 30 cm H(2)O), and high (VT, 1,000 mL; PIP, 35 cm H(2)O) ventilator settings.
When the ventilator low-inspiratory-pressure alarm was set at 4 cm H(2)O below the desired PIP, it failed to alarm for simulated decannulation of tracheostomy tubes < 4.5 mm on low and medium settings, and < 4.0 mm on high settings. When the ventilator low-inspiratory-pressure alarm was set at 10 cm H(2)O below the desired PIP, it failed to alarm with tracheostomy tubes < 6.0 mm.
We conclude that ventilator low-inspiratory-pressure alarms fail to alarm during simulated decannulation with small tracheostomy tubes commonly used in children. We speculate that low-inspiratory-pressure alarms set at 4 cm H(2)O below the desired PIP will detect more decannulation than when set at 10 cm H(2)O below the desired PIP.
正压通气机配备了低吸气压力警报器,以保护患者避免通气不足。无套囊的小儿气管造口插管阻力较高,在拔管期间可能不会触发这些警报。
确定通气机低吸气压力警报器在检测气管造口拔管方面是否有效。
我们将不同内径(3.0至6.0毫米)的气管造口插管连接到家用通气机上,并使用低(潮气量[VT],600毫升;吸气峰压[PIP],25厘米水柱)、中(VT,800毫升;PIP,30厘米水柱)和高(VT,1000毫升;PIP,35厘米水柱)通气机设置模拟拔管。
当通气机低吸气压力警报器设置为比预期PIP低4厘米水柱时,对于内径<4.5毫米的气管造口插管,在低和中设置下模拟拔管时未发出警报,在高设置下内径<4.0毫米时未发出警报。当通气机低吸气压力警报器设置为比预期PIP低10厘米水柱时,对于内径<6.0毫米的气管造口插管未发出警报。
我们得出结论,在使用儿童常用的小气管造口插管进行模拟拔管期间,通气机低吸气压力警报器未发出警报。我们推测,设置为比预期PIP低4厘米水柱的低吸气压力警报器比设置为比预期PIP低10厘米水柱时能检测到更多的拔管情况。