Singh A K, McArdle C, Gordon B, Ashraf M, Granley K
Department of Veterinary Diagnostic Investigation, College of Veterinary Medicine, University of Minnesota, St. Paul.
Biomed Chromatogr. 1989 Nov;3(6):262-5. doi: 10.1002/bmc.1130030607.
A high performance liquid chromatographic method is described for the simultaneous determination of furosemide and bumetanide in horse plasma. The C8 (3 microns) reversed phase column (4.8 x 150 mm) provided clear separation of furosemide and bumetanide with other components present in the horse plasma. The detection limit for both the drugs was 10 ng/mL. Both drugs were stable in plasma (at natural or acidic pH) for up to 24 h. The method is sufficiently sensitive to detect furosemide levels in plasma obtained from horses receiving a therapeutic dose of furosemide.