Adverse local tissue reactions in metal-on-polyethylene total hip arthroplasty due to trunnion corrosion: the risk of misdiagnosis.

作者信息

Whitehouse M R, Endo M, Zachara S, Nielsen T O, Greidanus N V, Masri B A, Garbuz D S, Duncan C P

机构信息

University of Bristol, Bristol, BS10 5NB, UK.

Kyushu University Hospital, 3-1-1, Maidashi, Higashi-ku, Fukuoka City, 812-8582, Japan.

出版信息

Bone Joint J. 2015 Aug;97-B(8):1024-30. doi: 10.1302/0301-620X.97B8.34682.

Abstract

Adverse reaction to wear and corrosion debris is a cause for concern in total hip arthroplasty (THA). Modular junctions are a potential source of such wear products and are associated with secondary pseudotumour formation. We present a consecutive series of 17 patients treated at our unit for this complication following metal-on-highly cross-linked polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as infection, and present the aggregate laboratory results and pathological findings in this series. The clinical presentation was pain, swelling or instability. Solid, cystic and mixed soft-tissue lesions were noted on imaging and confirmed intra-operatively. Corrosion at the head-neck junction was noted in all cases. No bacteria were isolated on multiple pre- and intra-operative samples yet the mean erythrocyte sedimentation rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and stromal polymorphonuclear cell counts were noted in nine cases. Adverse soft-tissue reactions can occur in MoP THA owing to corrosion products released from the head-neck junction. The diagnosis should be carefully considered when investigating pain after THA. This may avoid the misdiagnosis of periprosthetic infection with an unidentified organism and mitigate the unnecessary management of these cases with complete single- or two-stage exchange.

摘要

全髋关节置换术(THA)中,对磨损及腐蚀碎屑的不良反应令人担忧。模块化连接部位是此类磨损产物的一个潜在来源,且与继发性假肿瘤形成有关。我们报告了在本单位接受治疗的连续17例因金属对高交联聚乙烯(MoP)全髋关节置换术后出现该并发症的患者。我们强调了误诊为感染的风险,并展示了该系列患者的综合实验室检查结果及病理发现。临床表现为疼痛、肿胀或不稳定。影像学检查发现了实性、囊性及混合性软组织病变,并在术中得到证实。所有病例均发现头颈交界处存在腐蚀。多次术前及术中样本均未分离出细菌,但平均红细胞沉降率为49(9至100),C反应蛋白为32(0.6至106),9例患者出现基质多形核细胞计数。由于头颈交界处释放的腐蚀产物,MoP全髋关节置换术中可能会发生不良软组织反应。在调查全髋关节置换术后疼痛时,应仔细考虑该诊断。这可以避免将假体周围感染误诊为不明病原体感染,并减少对这些病例进行不必要的单期或两期完全翻修治疗。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验