恰加斯病中的变时性功能不全:混合传感器(每分钟容积和加速度计)的有效性。

Chronotropic incompetence in Chagas disease: effectiveness of blended sensor (volume/minute and accelerometer).

作者信息

Menezes Junior Antonio da Silva, Silva Aline Pereira da, Profahl Giovana Gurian Batista, Ottobeli Catarine, Louzeiro Jutay Fernando Silva

机构信息

Pontifícia Universidade Católica de Goiás, Goiânia, GO, BR.

出版信息

Rev Bras Cir Cardiovasc. 2015 Jul-Sep;30(3):311-5. doi: 10.5935/1678-9741.20150035.

Abstract

INTRODUCTION

Technological progress of pacemakers has allowed the association of two or more sensors in one heart rate system response. The accelerometer sensor measures the intensity of the activity; it has a relatively rapid response to the beginning of it, however, it may present insufficient response to less strenuous or of less impact exercise. The minute ventilation sensor changes the pacing rate in response to changes in respiratory frequency in relation to tidal volume, allowing responses to situations of emotional stress and low impact exercises.

OBJECTIVE

To evaluate the cardiorespiratory response of the accelerometer with respect to the blended sensor (BS=accelerometer sensor+minute ventilation sensor) to exercise in chagasic patients undergoing cardiopulmonary exercise test.

METHODS

This was a prospective, observational, randomized, cross-sectional study. Patients who met the inclusion criteria were selected. The maximum heart rate of the sensor was programmed by age (220-age). The results were analyzed through t test with paired samples (P<0.05).

RESULTS

Sample was comprised of 44 patients, with a mean age of 66±10.4 years, 58% were female, 54% as first implant, in 74% were functional class I and 26% were functional class II, left ventricular ejection fraction was 58±7. As for the cardiopulmonary test, maximum expected heart rate and VO2 were not achieved in both the accelerometer sensor and the blended sensor, however, metabolic equivalent in the blended sensor was higher than the expected, all data with P<0.001.

CONCLUSION

Even though the maximal heart rate was not reached, the blended sensor provided a physiological electrical sequence when compared to the accelerometer sensor, providing better physical fitness test in cardiopulmonary hemodynamics and greater efficiency.

摘要

引言

起搏器的技术进步使得在一个心率系统响应中能够结合两个或更多传感器。加速度计传感器测量活动强度;它对活动开始有相对快速的响应,然而,对于强度较小或冲击力较小的运动,它可能表现出响应不足。分钟通气量传感器根据呼吸频率相对于潮气量的变化来改变起搏频率,从而对情绪应激和低冲击力运动情况做出响应。

目的

在接受心肺运动试验的恰加斯病患者中,评估加速度计相对于混合传感器(BS = 加速度计传感器 + 分钟通气量传感器)对运动的心肺反应。

方法

这是一项前瞻性、观察性、随机、横断面研究。选择符合纳入标准的患者。根据年龄(220 - 年龄)对传感器的最大心率进行编程。通过配对样本t检验分析结果(P < 0.05)。

结果

样本包括44名患者,平均年龄为66±10.4岁,58%为女性,54%为首次植入,74%为心功能I级,26%为心功能II级,左心室射血分数为58±7。至于心肺试验,加速度计传感器和混合传感器均未达到最大预期心率和VO2,然而,混合传感器中的代谢当量高于预期,所有数据P < 0.001。

结论

尽管未达到最大心率,但与加速度计传感器相比,混合传感器提供了生理电序列,在心肺血流动力学方面提供了更好的体能测试且效率更高。