Jia Yulin, Zhang Jinling, Fan Jinyu, Li Chao, Sun Yuanyuan, Li Di, Xiao Xigang
1 Department of CT, The First Affiliated Hospital of Harbin Medical University, Harbin, China.
2 Department of Medical Imaging, Heilongjiang Provincial Hospital, Harbin, China.
Br J Radiol. 2015;88(1055):20150222. doi: 10.1259/bjr.20150222. Epub 2015 Sep 2.
This study aimed to evaluate the effect of gemstone spectral imaging (GSI) for metal artefact reduction in cerebral artery CT angiography (CTA) after metal coils or clips treatment.
35 patients with cerebral aneurysms were treated with metal coils or clips and underwent CTA using gemstone spectral CT between February and December 2013. The data were reconstructed into three image groups including Group A (quality check images with 140 kVp), Group B (monochromatic image sets in the range of 40-140 keV) and Group C [monochromatic image sets with metal artefacts reduction software (MARS GE Medical Systems, Waukesha, WI)]. CT attenuation value of cerebral artery, contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR) and the subjective score of all images were measured and compared statistically.
CT attenuation value of cerebral artery decreased in Groups B and C as the photon energy increased. The average energy levels of 60.05 ± 5.37 and 59.93 ± 5.57 keV presented the best CNR in Groups B and C, respectively. CNR values, SNR values and the subjective scores of the image quality of the two sets were higher than those of Group A.
GSI reduced metal artefact and improved the image quality of CTA after metal coils or clips treatment in patients with cerebral aneurysm. The monochromatic images at the average energy level of 60.05 ± 5.37 keV with MARS and 59.93 ± 5.57 keV without MARS were suggested to be the optimal parameters.
GSI could reduce metal artefact after metal coils or clips treatment in patients with cerebral aneurysm.
本研究旨在评估宝石光谱成像(GSI)对金属线圈或夹子治疗后脑动脉CT血管造影(CTA)中金属伪影减少的效果。
35例脑动脉瘤患者接受了金属线圈或夹子治疗,并于2013年2月至12月间使用宝石光谱CT进行CTA检查。数据被重建为三个图像组,包括A组(140 kVp的质量检查图像)、B组(40 - 140 keV范围内的单色图像集)和C组[带有金属伪影减少软件(MARS GE Medical Systems,沃基沙,威斯康星州)的单色图像集]。测量并统计比较脑动脉的CT衰减值、对比噪声比(CNR)、信噪比(SNR)以及所有图像的主观评分。
随着光子能量增加,B组和C组脑动脉的CT衰减值降低。B组和C组中平均能量水平分别为60.05±5.37 keV和59.93±5.57 keV时呈现出最佳的CNR。两组的CNR值、SNR值和图像质量主观评分均高于A组。
GSI减少了金属伪影,并改善了脑动脉瘤患者金属线圈或夹子治疗后脑动脉CTA的图像质量。建议将使用MARS时平均能量水平为60.05±5.37 keV以及未使用MARS时平均能量水平为59.93±5.57 keV的单色图像作为最佳参数。
GSI可减少脑动脉瘤患者金属线圈或夹子治疗后的金属伪影。