Premachandra B N
J Clin Endocrinol Metab. 1978 Oct;47(4):746-50. doi: 10.1210/jcem-47-4-746.
rT3 RIA standard curves were generated using rT3 standards from various sources. The rT3 antiserum and the method of free and bound hormone separation used were the same in all assays. The slopes of rT3 RIA curves differed considerably; standard curves generated with Henning (Berlin) rT3 standard had the steepest slope. Because of the differences in slopes of standard curves, bound to free estimates, when extrapolated from standard curves generated with various rT3 standards, resulted in 2- to 3-fold differences in rT3 assay values in the same sera. Circulating rT3 concentration in human sera was 20 ng/100 ml when Henning rT3 standard was used in the assay, whereas with Jorgensen's rT3 standard, a value of 42 ng/100 ml was noted. As the rT3 antiserum and assay conditions used were identical in all comparisons (except rT3 standards), the 2- to 3-fold differences in rT3 values noted in the same sera suggested variability in potency of rT3 standards. The differences in rT3 values in the same sera (resulting from extrapolation from standard curves using different rT3 standards) were of the same magnitude as that noted in normal human sera in various studies, suggesting that the reported rT3 variations in sera of euthyroid individuals may also be due to differences in rT3 standards employed in the assay.
使用来自不同来源的反三碘甲状腺原氨酸(rT3)标准品生成rT3放射免疫分析(RIA)标准曲线。在所有分析中,所使用的rT3抗血清以及游离和结合激素分离方法均相同。rT3 RIA曲线的斜率差异很大;用亨宁(柏林)rT3标准品生成的标准曲线斜率最陡。由于标准曲线斜率的差异,当从用不同rT3标准品生成的标准曲线外推结合与游离估计值时,同一血清中rT3分析值会出现2至3倍的差异。当分析中使用亨宁rT3标准品时,人血清中循环rT3浓度为20 ng/100 ml,而使用约根森rT3标准品时,该值为42 ng/100 ml。由于在所有比较中(除rT3标准品外)所使用的rT3抗血清和分析条件相同,同一血清中rT3值出现2至3倍的差异表明rT3标准品的效价存在变异性。同一血清中rT3值的差异(由使用不同rT3标准品从标准曲线外推得出)与各项研究中正常人血清中所观察到的差异幅度相同,这表明在甲状腺功能正常个体血清中所报道的rT3变化也可能是由于分析中所使用的rT3标准品不同所致。