Simeon V
Arh Hig Rada Toksikol. 1989 Jun;40(2):183-9.
The commercially available kits (Test-Combination, Cholinesterase, Boehringer, Mannheim and Test-reagent, Cholinesterase EC 3.1.1.8, Pliva, Zagreb) and reagents prepared in own laboratory for measuring cholinesterase activity (Ellman method) were tested with respect to their stability and the reproducibility of the activity measurements. The reagents of the three sources were shown to be interchangeable and equally stable over a few weeks. The coefficient of variation for within-run measurements by the Ellman method was 2-3% and that for between-run measurements 6%. The stability of the few enzyme standards (Precinorm E, Precinorm U, NBS-serum and native human serum) for the quality control of the measurements was also tested: the most stable was native human serum.
对市售试剂盒(Test-Combination、Cholinesterase,Boehringer,Mannheim公司)、试剂(Test-reagent,Cholinesterase EC 3.1.1.8,Pliva,萨格勒布公司)以及在本实验室自行配制用于测定胆碱酯酶活性的试剂(Ellman法),就其稳定性和活性测量的重现性进行了测试。结果表明,这三种来源的试剂可互换使用,且在几周内具有同等稳定性。采用Ellman法进行批内测量的变异系数为2% - 3%,批间测量的变异系数为6%。还对用于测量质量控制的几种酶标准品(Precinorm E、Precinorm U、NBS血清和天然人血清)的稳定性进行了测试:最稳定的是天然人血清。