Braun Benedikt Johannes, Veith Nils Thomas, Hell Rebecca, Döbele Stefan, Roland Michael, Rollmann Mika, Holstein Jörg, Pohlemann Tim
Department of Trauma, Hand and Reconstructive Surgery, Saarland University, Building 57, Kirrbergerstr. 1, 66421 Homburg, Germany.
BG Trauma Center, Department of Trauma Surgery, Eberhard Karls University Tübingen, Tübingen, Germany.
J Foot Ankle Res. 2015 Sep 22;8:54. doi: 10.1186/s13047-015-0111-8. eCollection 2015.
A new tool (OpenGo, Moticon GmbH) was introduced to continuously measure kinetic and temporospatial gait parameters independently through an insole over up to 4 weeks. The goal of this study was to investigate the validity and reliability of this new insole system in a group of healthy individuals.
Gait data were collected from 12 healthy individuals on a treadmill at two different speeds. In total, six trials of three minutes each were performed by every participant. Validation was performed with the FDM-S System (Zebris). Complete sensor data were used for a within test reliability analysis of over 10000 steps. Intraclass correlation was calculated for different gait parameters and analysis of variance performed.
Intraclass correlation for the validation was >0.796 for temporospatial and kinetic gait parameters. No statistical difference was seen between the insole and force plate measurements (difference between means: 36.3 ± 27.19 N; p = 0.19 and 0.027 ± 0.028 s; p = 0.36). Intraclass correlation for the reliability was >0.994 for all parameters measured.
The system is feasible for clinical trials that require step by step as well as grouped analysis of gait over a long period of time. Comparable validity and reliability to a stationary analysis tool has been shown.
引入了一种新工具(OpenGo,Moticon GmbH公司),可通过鞋垫连续独立测量长达4周的动力学和时空步态参数。本研究的目的是调查这种新型鞋垫系统在一组健康个体中的有效性和可靠性。
从12名健康个体在跑步机上以两种不同速度收集步态数据。每位参与者总共进行了6次每次3分钟的试验。使用FDM-S系统(Zebris)进行验证。完整的传感器数据用于超过10000步的测试内可靠性分析。计算不同步态参数的组内相关性并进行方差分析。
时空和动力学步态参数验证的组内相关性>0.796。鞋垫测量和测力板测量之间未见统计学差异(均值差异:36.3±27.19 N;p = 0.19和0.027±0.028 s;p = 0.36)。所有测量参数的可靠性组内相关性>0.994。
该系统对于需要长时间逐步步态分析以及分组步态分析的临床试验是可行的。已显示出与静态分析工具相当的有效性和可靠性。