Liao Xiongfei, Wang Yunlai, Lang Jinyi, Wang Pei, Li Jie, Ge Ruigang, Yang Jack
Department of Radiation Oncology, Sichuan Cancer Hospital, 55 4th Section Renmin South Road, Chengdu, 610041, China.
Department of Radiation Oncology, Chinese PLA General Hospital, 28 Fuxing Avenue, Beijing, 100853, China.
Biomed Mater Eng. 2015;26 Suppl 1:S1659-67. doi: 10.3233/BME-151465.
To evaluate the Elekta kilovoltage CBCT doses and the associated technical protocols with patient dosimetry estimation. Image guidance technique with cone-beam CT (CBCT) in radiation oncology on a daily basis can deliver a significant dose to the patient. To evaluate the patient dose from LINAC-integrated kV cone beam CT imaging in image-guided radiotherapy. CT dose index (CTDI) were measured with PTW TM30009 CT ion chamber in air, in head phantom and body phantom, respectively; with different combinations of tube voltage, current, exposure time per frame, collimator and gantry rotation range. Dose length products (DLP) were subsequently calculated to account for volume integration effects. The CTDI and DLP were also compared to AcQSim™ simulator CT for routine clinical protocols. Both CTDIair and CTDIw depended quadratically on the voltage, while linearly on milliampere x seconds (mAs) settings. It was shown that CTDIw and DLP had very close relationship with the collimator settings and the gantry rotation ranges. Normalized CTDIw for Elekta XVI™ CBCT was lower than that of ACQSim simulator CT owing to its pulsed radiation output characteristics. CTDIw can be used to assess the patient dose in CBCT due to its simplicity for measurement and reproducibility. Regular measurement should be performed in QA & QC program. Optimal image parameters should be chosen to reduce patient dose during CBCT.
评估医科达千伏锥形束CT剂量以及相关技术方案,并进行患者剂量测定估算。放射肿瘤学中每日使用的锥形束CT(CBCT)图像引导技术会给患者带来显著剂量。为评估图像引导放射治疗中直线加速器集成千伏锥形束CT成像的患者剂量。分别在空气中、头部模体和体部模体中,使用PTW TM30009 CT电离室,在不同管电压、电流、每帧曝光时间、准直器和机架旋转范围的组合下测量CT剂量指数(CTDI);随后计算剂量长度乘积(DLP)以考虑体积积分效应。还将CTDI和DLP与用于常规临床方案的AcQSim™模拟机CT进行比较。CTDIair和CTDIw均与电压呈二次方关系,而与毫安秒(mAs)设置呈线性关系。结果表明,CTDIw和DLP与准直器设置和机架旋转范围有非常密切的关系。由于其脉冲辐射输出特性,医科达XVI™ CBCT的归一化CTDIw低于ACQSim模拟机CT。CTDIw因其测量简单且具有可重复性,可用于评估CBCT中的患者剂量。在质量保证和质量控制程序中应定期进行测量。应选择最佳图像参数以减少CBCT期间的患者剂量。