Abstract

On-line megavoltage computed tomography (MVCT) images are used to verify patient daily set-up in Hi-ART helical TomoTherapy unit. To evaluate the patient dose from MVCT scanning in image guidance, weighted computed tomography (CT) dose index (CTDI) was measured with PTW TM30009 CT pencil chamber in head and body phantoms for slice thicknesses of 2, 4 and 6 mm with different scan lengths. Dose length products (DLPs) were subsequently calculated. The CTDI and DLP were compared with XVI kV CBCT and Brilliance simulator CT for routine clinical protocols. It was shown that CTDI and DLP had close relationship with the slice thickness and the scan length. The dose distribution in the transversal plane was very inhomogeneous due to the attenuation of the couch. Patient dose from MVCT was lower than XVI CBCT for the head scan, while larger for body scan. CTDI, which is measured easily and reproducibly, can be used to assess the patient dose in MVCT. Regular measurement should be performed in QA & QC programmes. Appropriate slice thickness and scan range should be chosen to reduce the patient dose.

摘要

在线兆伏级计算机断层扫描(MVCT)图像用于在螺旋断层放疗系统中验证患者的每日摆位。为评估图像引导中MVCT扫描的患者剂量,使用PTW TM30009 CT铅笔电离室在头部和体模中测量了不同扫描长度下2、4和6毫米层厚的加权计算机断层扫描(CT)剂量指数(CTDI)。随后计算了剂量长度乘积(DLP)。将CTDI和DLP与用于常规临床方案的XVI kV CBCT和飞利浦模拟定位机CT进行了比较。结果表明,CTDI和DLP与层厚和扫描长度密切相关。由于治疗床的衰减,横断面的剂量分布非常不均匀。头部扫描时,MVCT的患者剂量低于XVI CBCT,而体部扫描时则更高。CTDI测量简便且可重复,可用于评估MVCT中的患者剂量。在质量保证和质量控制程序中应定期进行测量。应选择合适的层厚和扫描范围以降低患者剂量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验