Moncada Jeanne, Shayevich Clara, Philip Susan S, Lucic Danijela, Schachter Julius
From the *Department of Laboratory Medicine, University of California, San Francisco, CA; †San Francisco Department of Public Health, San Francisco, CA; and ‡Abbott Molecular Inc, Des Plaines, IL.
Sex Transm Dis. 2015 Nov;42(11):650-1. doi: 10.1097/OLQ.0000000000000360.
We evaluated Abbott's RealTime assay for the detection of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) in the urethra, oropharynx, and rectum of 260 men who have sex with men. Compared with Hologic's AC2, RealTime had good agreement for detecting CT and GC. Overall, there were 25 CT and 44 GC AC2 positives, and 26 CT and 38 GC RealTime positives. For total negatives, there were 742 CT and 725 GC for AC2, 744 CT and 724 GC for RealTime.
我们评估了雅培实时检测法在260名男男性行为者的尿道、口咽和直肠中检测沙眼衣原体(CT)和淋病奈瑟菌(GC)的情况。与豪洛捷公司的AC2检测法相比,实时检测法在检测CT和GC方面具有良好的一致性。总体而言,AC2检测出25例CT阳性和44例GC阳性,实时检测法检测出26例CT阳性和38例GC阳性。对于全部阴性结果,AC2检测出742例CT阴性和725例GC阴性,实时检测法检测出744例CT阴性和724例GC阴性。