Faculty of Science and Information Technology, School of Mathematical and Physical Sciences, University of Newcastle, New South Wales, Australia.
Faculty of Engineering and Built Environment, School of Electrical Engineering and Computer Science, University of Newcastle, New South Wales, Australia.
Int J Radiat Oncol Biol Phys. 2015 Nov 1;93(3):516-22. doi: 10.1016/j.ijrobp.2015.07.2271. Epub 2015 Jul 22.
Gantry-mounted megavoltage electronic portal imaging devices (EPIDs) have become ubiquitous on linear accelerators. WatchDog is a novel application of EPIDs, in which the image frames acquired during treatment are used to monitor treatment delivery in real time. We report on the preliminary use of WatchDog in a prospective study of cancer patients undergoing intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) and identify the challenges of clinical adoption.
At the time of submission, 28 cancer patients (head and neck, pelvis, and prostate) undergoing fractionated external beam radiation therapy (24 IMRT, 4 VMAT) had ≥1 treatment fraction verified in real time (131 fractions or 881 fields). EPID images acquired continuously during treatment were synchronized and compared with model-generated transit EPID images within a frame time (∼0.1 s). A χ comparison was performed to cumulative frames to gauge the overall delivery quality, and the resulting pass rates were reported graphically during treatment delivery. Every frame acquired (500-1500 per fraction) was saved for postprocessing and analysis.
The system reported the mean ± standard deviation in real time χ 91.1% ± 11.5% (83.6% ± 13.2%) for cumulative frame χ analysis with 4%, 4 mm (3%, 3 mm) criteria, global over the integrated image.
A real-time EPID-based radiation delivery verification system for IMRT and VMAT has been demonstrated that aims to prevent major mistreatments in radiation therapy.
龙门架式兆伏级电子射野影像装置(EPID)已在直线加速器上得到广泛应用。WatchDog 是 EPID 的一种新应用,它利用治疗过程中采集的图像帧实时监测治疗的实施情况。我们报告了在一项针对接受强度调制放射治疗(IMRT)和容积调制弧形治疗(VMAT)的癌症患者的前瞻性研究中使用 WatchDog 的初步结果,并确定了临床应用的挑战。
在提交时,28 名癌症患者(头颈部、骨盆和前列腺)正在接受分次外照射放射治疗(24 例 IMRT,4 例 VMAT),其中至少有 1 个治疗分次实时验证(131 个分次或 881 个射野)。在治疗过程中连续采集的 EPID 图像与框架时间内(约 0.1 s)生成的模型传输 EPID 图像进行同步和比较。通过 χ 比较对累积帧进行评估,以衡量整体输送质量,并用图形在治疗过程中报告生成的通过率。每帧采集(每个分次 500-1500 帧)都用于后处理和分析。
系统实时报告 χ 平均值 ± 标准偏差为 91.1% ± 11.5%(累积帧 χ 分析的 83.6% ± 13.2%),4%,4 毫米(3%,3 毫米)的标准,全局图像上的。
已经证明了一种用于 IMRT 和 VMAT 的基于实时 EPID 的放射治疗验证系统,旨在防止放射治疗中的重大失误。