人工晶状体植入术后预测残余散光方法的比较

Comparison of Methods to Predict Residual Astigmatism After Intraocular Lens Implantation.

作者信息

Abulafia Adi, Hill Warren E, Franchina Maria, Barrett Graham D

出版信息

J Refract Surg. 2015 Oct;31(10):699-707. doi: 10.3928/1081597X-20150928-03.

Abstract

PURPOSE

To evaluate and compare the accuracy of three toric intraocular lens (IOL) calculators using keratometry measurements derived from the anterior corneal curvature and direct measurements of the posterior corneal curvature.

METHODS

Postoperative corneal astigmatism was measured by the IOLMaster (Carl Zeiss Meditec, Jena, Germany) and Pentacam (Oculus Optikgeräte, Wetzlar, Germany). The data were processed by the Alcon, Holladay, and Barrett toric IOL calculators. The error in predicted residual astigmatism (PredRA) was calculated by subtracting the PredRA from the postoperative subjective refraction by vector analysis.

RESULTS

The centroid errors in PredRA were against-the-rule (ATR) with the Alcon (0.56 diopters [D]) and Holladay (0.55 D) toric calculators using the IOLMaster (Carl Zeiss Meditec, Jena, Germany) measurements. The centroid errors in PredRA were lower when Pentacam (Oculus Optikgeräte, Wetzlar, Germany) measurements were used (0.38 D, ATR). The Barrett toric calculator using the IOLMaster measurements had the lowest centroid errors in PredRA (0.02 D, P < .001) and achieved the most accurate results: 75.8% and 92.9% of eyes were within 0.50 and 0.75 D of the PredRA, respectively.

CONCLUSIONS

The prediction of the postoperative astigmatic outcome can be improved by using appropriate methods of adjustment for posterior corneal astigmatism.

摘要

目的

使用源自角膜前表面曲率的角膜曲率测量值和角膜后表面曲率的直接测量值,评估并比较三种散光人工晶状体(IOL)计算器的准确性。

方法

使用IOLMaster(德国耶拿卡尔蔡司医疗技术公司)和Pentacam(德国韦茨拉尔奥culus光学设备公司)测量术后角膜散光。数据由爱尔康、霍拉迪和巴雷特散光IOL计算器进行处理。通过矢量分析,用术后主观验光值减去预测残余散光(PredRA)来计算预测残余散光的误差。

结果

使用IOLMaster(德国耶拿卡尔蔡司医疗技术公司)测量值时,爱尔康(0.56屈光度[D])和霍拉迪(0.55 D)散光计算器的PredRA质心误差为逆规散光(ATR)。使用Pentacam(德国韦茨拉尔奥culus光学设备公司)测量值时,PredRA的质心误差较低(0.38 D,逆规散光)。使用IOLMaster测量值的巴雷特散光计算器的PredRA质心误差最低(0.02 D,P <.001),并获得了最准确的结果:分别有75.8%和92.9%的眼睛的PredRA在0.50和0.75 D范围内。

结论

通过使用适当的方法调整角膜后表面散光,可以改善术后散光结果的预测。