Thyroid radiation dose during panoramic and cephalometric dental x-ray examinations.

作者信息

Block A J, Goepp R A, Mason E W

出版信息

Angle Orthod. 1977 Jan;47(1):17-24. doi: 10.1043/0003-3219(1977)047<0017:TRDDPA>2.0.CO;2.

Abstract

Radiation exposure from panoramic equipment can be reduced significantly by use of smaller film, adjustment of the beam height to the height of the smaller film, and careful positioning of patients. These techniques have no adverse effect on the quality of the diagnostic information needed in dentistry. In addition to describing methods of reducing exposures from panoramic machines, this study demonstrates that the use of a barrier collar during static, cephalometric examinations can appreciably reduce thyroid exposure. Since the objective is to obtain diagnostic information from the film without irradiating the thyroid, the application of a lead-impregnated collar is a minor inconvenience, easily borne by the patient and operator. It should be noted that the use of the collar during panoramic examinations affords little or no protection since the relative motion of the panoramic machine places the axis of movement inside the head and neck of the patient. While the evolution of diagnostic radiology may have reached a high level of technical refinement of equipment and film the clinician still must avoid unnecessary exposure for X-ray examinations and must carefully select the best type of examination to be used for each patient. For example, a complete panoramic examination to determine the position of a known unerupted third molar tooth is probably not an exercise of good judgment since other examinations, such as periapical, could yield the same information with less exposure. Decisions must be made with good judgment, value being placed on relative risks versus the benefits of diagnostic yield.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验