使用光学相干断层扫描技术对预期与实际的Intacs角膜基质内环形片进行深度评估。

Depth evaluation of intended vs actual intacs intrastromal ring segments using optical coherence tomography.

作者信息

Barbara R, Barbara A, Naftali M

机构信息

Department of Ophthalmology, Bnai Zion Medical Centre, Haifa, Israel.

Hadassah Optimal Private Medical Centre, Affiliated to Hadassah Medical Centre, Haifa, Israel.

出版信息

Eye (Lond). 2016 Jan;30(1):102-10. doi: 10.1038/eye.2015.202. Epub 2015 Oct 23.

Abstract

PurposeEvaluation of actual vs intended intrastromal corneal ring segments (ICRS) implantation depth as measured by anterior segment optical coherence tomography (OCT)MethodsProspective study evaluating 30 Intacs segments implanted manually in 19 eyes of 15 patients suffering from keratoconus. Segment depth evaluation was performed using anterior segment OCT. Measurements were performed above and below the segment at 3 points in relation to the incision site. Statistical analysis was performed using the SAS software for ANOVA, matched t-test, and GLIMMIX procedure.ResultsIntacs segment depth was 153-μm shallower than intended (58% vs 80%). Segment layout demonstrated the proximal and distal portions to be 13-μm shallower and 12-μm deeper (on average), respectively. Intacs segment thickness does not influence implantation depth. Intacs segments implanted in the same eye do not share similar implantation depths. Stromal compression is likely to occur.ConclusionIntacs are implanted at a shallower depth than intended. The 'pocketing' stage prior to implantation most likely has a stronger effect on the segment's final implantation depth than does the incisions' depth.

摘要

目的

通过眼前节光学相干断层扫描(OCT)评估基质内角膜环片(ICRS)实际植入深度与预期植入深度的差异

方法

前瞻性研究,对15例圆锥角膜患者的19只眼手动植入30个Intacs环片。使用眼前节OCT进行环片深度评估。在与切口部位相关的3个点处,于环片上方和下方进行测量。使用SAS软件进行方差分析、配对t检验和广义线性混合模型(GLIMMIX)程序进行统计分析。

结果

Intacs环片深度比预期浅153μm(58%对80%)。环片布局显示近端和远端部分分别平均浅13μm和深12μm。Intacs环片厚度不影响植入深度。同一眼内植入的Intacs环片不具有相似的植入深度。可能会发生基质压缩。

结论

Intacs环片植入深度比预期浅。植入前的“成袋”阶段对环片最终植入深度的影响可能比切口深度的影响更大。

相似文献

1
Depth evaluation of intended vs actual intacs intrastromal ring segments using optical coherence tomography.
Eye (Lond). 2016 Jan;30(1):102-10. doi: 10.1038/eye.2015.202. Epub 2015 Oct 23.
2
Optical coherence tomography to assess intrastromal corneal ring segment depth in keratoconic eyes.
J Cataract Refract Surg. 2006 Nov;32(11):1860-5. doi: 10.1016/j.jcrs.2006.05.030.
3
Depth of intrastromal corneal ring segments by OCT.
Eur J Ophthalmol. 2013 Mar-Apr;23(2):171-6. doi: 10.5301/ejo.5000212.
5
Intraoperative use of AS-OCT during intrastromal corneal ring segment implantation.
Ophthalmic Surg Lasers Imaging. 2012 Nov-Dec;43(6 Suppl):S109-16. doi: 10.3928/15428877-20121003-04.
6
Assessment of intrastromal corneal ring segment position with anterior segment optical coherence tomography.
Ophthalmic Surg Lasers Imaging. 2012 May-Jun;43(3):214-21. doi: 10.3928/15428877-20120301-01. Epub 2012 Mar 8.
7
Measurement of depth of Intacs implanted via femtosecond laser using Pentacam.
J Refract Surg. 2009 Apr;25(4):377-82. doi: 10.3928/1081597X-20090401-05.
8
Experimental imaging of intracorneal ring segments with optical coherence tomography.
J Refract Surg. 2003 May-Jun;19(3):367-71. doi: 10.3928/1081-597X-20030501-16.
9
INTACS followed by KeraRing intrastromal corneal ring segment implantation for keratoconus.
J Refract Surg. 2010 May;26(5):371-4. doi: 10.3928/1081597X-20100203-01.
10

引用本文的文献

1
Delayed Perforation of an Intrastromal Corneal Ring Segment into the Anterior Chamber: A Case Report and Review of the Literature.
Case Rep Ophthalmol. 2021 Sep 9;12(3):740-748. doi: 10.1159/000518012. eCollection 2021 Sep-Dec.
2
Corneal Biomechanics After Intrastromal Ring Surgery: Optomechanical In Silico Assessment.
Transl Vis Sci Technol. 2020 Oct 21;9(11):26. doi: 10.1167/tvst.9.11.26. eCollection 2020 Oct.
3
Optical coherence tomography for ocular surface and corneal diseases: a review.
Eye Vis (Lond). 2018 Jun 12;5:13. doi: 10.1186/s40662-018-0107-0. eCollection 2018.
5
6
Intracorneal ring segment depth in keratoconus patients: a long-term follow-up study.
Int Ophthalmol. 2018 Aug;38(4):1379-1383. doi: 10.1007/s10792-017-0595-2. Epub 2017 Jun 12.

本文引用的文献

1
Depth of intrastromal corneal ring segments by OCT.
Eur J Ophthalmol. 2013 Mar-Apr;23(2):171-6. doi: 10.5301/ejo.5000212.
2
Assessment of intrastromal corneal ring segment position with anterior segment optical coherence tomography.
Ophthalmic Surg Lasers Imaging. 2012 May-Jun;43(3):214-21. doi: 10.3928/15428877-20120301-01. Epub 2012 Mar 8.
4
Causes of intrastromal corneal ring segment explantation: clinicopathologic correlation analysis.
J Cataract Refract Surg. 2010 Jun;36(6):970-7. doi: 10.1016/j.jcrs.2009.12.042.
5
6
Measurement of depth of Intacs implanted via femtosecond laser using Pentacam.
J Refract Surg. 2009 Apr;25(4):377-82. doi: 10.3928/1081597X-20090401-05.
8
Intracorneal rings for keratoconus and keratectasia.
J Cataract Refract Surg. 2007 Jul;33(7):1303-14. doi: 10.1016/j.jcrs.2007.02.048.
9
Optical coherence tomography to assess intrastromal corneal ring segment depth in keratoconic eyes.
J Cataract Refract Surg. 2006 Nov;32(11):1860-5. doi: 10.1016/j.jcrs.2006.05.030.
10
European clinical evaluation: use of Intacs for the treatment of keratoconus.
J Cataract Refract Surg. 2006 May;32(5):747-55. doi: 10.1016/j.jcrs.2006.01.064.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验