Najafi Mohsen, Deevband Mohammad Reza, Ahmadi Mohsen, Kardan Mohammad Reza
Department of Bioengineering and Medical Physics, Shahid-Beheshti University of Medical Sciences, P. O. Box 14155-6446, Tehran, Iran.
Nuclear Sciences Research School, Nuclear Sciences and Technology Research Institute, P. O. Box 81745-313, Tehran, Iran.
Australas Phys Eng Sci Med. 2015 Dec;38(4):603-9. doi: 10.1007/s13246-015-0388-8. Epub 2015 Oct 27.
In order to establish diagnostic reference levels (DRLs) for multi-detector computed tomography (MDCT), four routine CT examinations were identified and a computer program was developed to collect data from 19 MDCT scanners in Iran. Mean values of Volume computed tomography dose index (CTDIvol) and dose-length product (DLP) in each site were calculated and the DRLs were defined as the 75th percentile of the distribution of the CTDIvol/DLP values for each examination. In terms of DLP, the DRLs of adult age group are 700, 290, 330, and 550 mGy cm for the Brain, Sinus, Chest, and Abdomen and Pelvis protocols, respectively. Although DRLs of this study are comparable to other international DRLs and in most cases are less than the international reference values, the great extent of dose distributions indicates that the CT imaging procedures in Iran should be optimized by applying diagnostic reference levels in order to decrease the radiation dose to patient undergoing CT examination.
为了建立多排螺旋计算机断层扫描(MDCT)的诊断参考水平(DRLs),确定了四项常规CT检查,并开发了一个计算机程序,从伊朗的19台MDCT扫描仪收集数据。计算了每个部位的容积计算机断层扫描剂量指数(CTDIvol)和剂量长度乘积(DLP)的平均值,并将DRLs定义为每次检查CTDIvol/DLP值分布的第75百分位数。就DLP而言,成人年龄组的脑部、鼻窦、胸部以及腹部和骨盆检查方案的DRLs分别为700、290、330和550 mGy cm。尽管本研究的DRLs与其他国际DRLs相当,且在大多数情况下低于国际参考值,但剂量分布范围较大表明,伊朗的CT成像程序应通过应用诊断参考水平进行优化,以降低接受CT检查患者的辐射剂量。