Zhang Jun-Peng, Liu Hua-Yan, Ning Chun-Ping, Chong Jing, Sun Yong-Mei
Department of Ultrasonography, the Affiliated Hospital of Qingdao University, Qingdao, China E-mail :
Asian Pac J Cancer Prev. 2015;16(16):7291-4. doi: 10.7314/apjcp.2015.16.16.7291.
PURPSOE: To investigate the diagnostic value of quantitative analysis of a tissue diffusion and virtual touch tissue imaging quantification (VTIQ) technique with acoustic radiation force impulse (ARFI) elastography for assessing enlarged cervical lymph nodes.
Fifty-six enlarged cervical lymph nodes confirmed by pathologic diagnoses were covered in the study. According to the results of pathologic diagnosis, patients were classified into benign and malignant groups. All the patients were examined by both conventional ultrasonography and elastography. AREA% and shear wave velocity (SWV) in ROI of different groups were calculated and compared using ROC curves. Cut-off points of AREA% and SWV were determined with receiver operating characteristic curves.
Final histopathological results revealed 21 cases of benign and 35 cases of malignant lymph nodes. The mean values of AREA% and SWV in benign and malignant groups were 45.0 ± 17.9% and 2.32 ± 0.57 m/s, and 61.3 ± 21.29% and 4.36 ± 1.25)m/s, respectively. For the parameters of elastography, "AREA%"and SWV demonstrated significant differences between groups (p=0.002). AREA% was positively correlated with SWV with a correlation coefficient of 0.809 (P<0.001).
Stiffness of different lymph node diseases in patients may differ. Elastography can evaluate changes sensitively and provide valuable information to doctors. The study proved that the VTIQ elastography technique can play an important role in differential diagnosis of lymph nodes.
目的:探讨组织弥散定量分析及虚拟触诊组织成像定量(VTIQ)技术联合声辐射力脉冲(ARFI)弹性成像在评估颈部肿大淋巴结中的诊断价值。
本研究纳入56例经病理诊断证实的颈部肿大淋巴结患者。根据病理诊断结果,将患者分为良性组和恶性组。所有患者均接受传统超声和弹性成像检查。计算不同组感兴趣区域(ROI)的面积百分比(AREA%)和剪切波速度(SWV),并采用ROC曲线进行比较。通过受试者操作特征曲线确定AREA%和SWV的截断点。
最终组织病理学结果显示,良性淋巴结21例,恶性淋巴结35例。良性组和恶性组AREA%的平均值分别为45.0±17.9%和2.32±0.57 m/s,SWV的平均值分别为61.3±21.29%和4.36±1.25 m/s。对于弹性成像参数,“AREA%”和SWV在两组之间存在显著差异(p = 0.002)。AREA%与SWV呈正相关,相关系数为0.809(P < 0.001)。
患者不同淋巴结疾病的硬度可能不同。弹性成像可敏感地评估变化并为医生提供有价值的信息。该研究证明VTIQ弹性成像技术在淋巴结的鉴别诊断中可发挥重要作用。