在老年人群中使用手持测力计获得的肌肉力量测量值的可靠性。

Reliability of muscle strength measures obtained with a hand-held dynamometer in an elderly population.

作者信息

Buckinx Fanny, Croisier Jean-Louis, Reginster Jean-Yves, Dardenne Nadia, Beaudart Charlotte, Slomian Justine, Leonard Sylvain, Bruyère Olivier

机构信息

Department of Public Health, Epidemiology and Health Economics, University of Liège, Liège, Belgium.

Support Unit in Epidemiology and Biostatistics, University of Liège, Liège, Belgium.

出版信息

Clin Physiol Funct Imaging. 2017 May;37(3):332-340. doi: 10.1111/cpf.12300. Epub 2015 Oct 30.

Abstract

BACKGROUND

The objective of this study was to assess the reliability of a hand-held dynamometer for isometric strength measurements among nursing home residents.

METHODS

The isometric muscle strength of nursing home residents was assessed for eight different muscle groups, using a hand-held dynamometer, the MicroFET2 device. Strength measurements were performed at baseline and after 4 days by the same operator and after 8 days by a second operator. Intraclass coefficients (ICC) were computed to assess the relative reliability, whereas the minimal detectable change (MDC%) was calculated to assess the absolute reliability of the test-retest of the MicroFET2 used by one single operator or by two different ones.

RESULTS

Thirty nursing home residents (75·0 ± 11·2 years, 50% of women) were enrolled in this study. ICC of the test-retest with one single operator ranged from 0·60 (0·37-0·83) for the ankle extensors to 0·85 (0·74-0·95) for the elbow flexors. When considering the test-retest with two different operators, the ICC values ranged from 0·62 (0·41-0·84) for the ankle extensors to 0·87 (0·79-0·96) for the elbow extensors. For the absolute reliability, MDC% varies from 27·64 (elbow flexors) to 81·97 (ankle extensors) when performed in intra-observer. In interobserver condition, MDC%, respectively, varies from 24·38 (elbow extensors) 67·59 (ankle extensors).

CONCLUSION

Using standardized protocol and standardized instructions to patients, a high relative and moderate absolute reliability was observed for all but ankle muscle groups, making this hand-held dynamometer a potential tool for research in the elderly population.

摘要

背景

本研究的目的是评估手持式测力计在测量养老院居民等长肌力时的可靠性。

方法

使用手持式测力计MicroFET2设备,对养老院居民的八个不同肌肉群的等长肌力进行评估。由同一名操作人员在基线时和4天后进行力量测量,8天后由第二名操作人员进行测量。计算组内相关系数(ICC)以评估相对可靠性,同时计算最小可检测变化(MDC%)以评估由一名操作人员或两名不同操作人员进行的MicroFET2重测的绝对可靠性。

结果

本研究纳入了30名养老院居民(75.0±11.2岁,50%为女性)。由一名操作人员进行的重测的ICC范围从踝背伸肌的0.60(0.37 - 0.83)到肘屈肌的0.85(0.74 - 0.95)。当考虑由两名不同操作人员进行的重测时,ICC值范围从踝背伸肌的0.62(0.41 - 0.84)到肘伸肌的0.87(0.79 - 0.96)。对于绝对可靠性,在同一观察者内进行测量时,MDC%从27.64(肘屈肌)到81.97(踝背伸肌)不等。在不同观察者条件下,MDC%分别从24.38(肘伸肌)到67.59(踝背伸肌)不等。

结论

对患者使用标准化方案和标准化指导语时,除踝部肌肉群外,所有肌肉群均观察到较高的相对可靠性和中等的绝对可靠性,这使得这种手持式测力计成为老年人群研究的潜在工具。