Department of Chemistry, North Carolina State University , Raleigh, North Carolina 27695, United States.
Promega Corporation, 2800 Woods Hollow Road, Madison, Wisconsin 53711, United States.
Anal Chem. 2015 Dec 1;87(23):11635-40. doi: 10.1021/acs.analchem.5b04121. Epub 2015 Nov 10.
We present a novel proteomic standard for assessing liquid chromatography-tandem mass spectrometry (LC-MS/MS) instrument performance, in terms of chromatographic reproducibility and dynamic range within a single LC-MS/MS injection. The peptide mixture standard consists of six peptides that were specifically synthesized to cover a wide range of hydrophobicities (grand average hydropathy (GRAVY) scores of -0.6 to 1.9). A combination of stable isotope labeled amino acids ((13)C and (15)N) were inserted to create five isotopologues. By combining these isotopologues at different ratios, they span four orders of magnitude within each distinct peptide sequence. Each peptide, from lightest to heaviest, increases in abundance by a factor of 10. We evaluate several metrics on our quadrupole orbitrap instrument using the 6 × 5 LC-MS/MS reference mixture spiked into a complex lysate background as a function of dynamic range, including mass measurement accuracy (MMA) and the linear range of quantitation of MS1 and parallel reaction monitoring experiments. Detection and linearity of the instrument routinely spanned three orders of magnitude across the gradient (500 fmol to 0.5 fmol on column) and no systematic trend was observed for MMA of targeted peptides as a function of abundance by analysis of variance analysis (p = 0.17). Detection and linearity of the fifth isotopologue (i.e., 0.05 fmol on column) was dependent on the peptide and instrument scan type (MS1 vs PRM). We foresee that this standard will serve as a powerful method to conduct both intra-instrument performance monitoring/evaluation, technology development, and inter-instrument comparisons.
我们提出了一种新的蛋白质组学标准,用于评估液相色谱-串联质谱(LC-MS/MS)仪器的性能,包括色谱重现性和单个 LC-MS/MS 注射的动态范围。肽混合物标准品由六种肽组成,这些肽是专门合成的,以涵盖广泛的疏水性(平均亲水性(GRAVY)评分范围为-0.6 至 1.9)。稳定同位素标记的氨基酸((13)C 和(15)N)被插入以创建五个同位素。通过以不同的比例组合这些同位素,它们在每个独特的肽序列中跨越四个数量级。每个肽,从轻到重,丰度增加 10 倍。我们使用 6×5 LC-MS/MS 参考混合物评估了我们的四极杆轨道阱仪器的几个指标,该混合物以复杂的裂解物背景中的浓度形式作为动态范围的函数,包括质量测量准确性(MMA)和 MS1 的定量线性范围平行反应监测实验。仪器的检测和线性度通常在梯度范围内跨越三个数量级(柱上 500 fmol 至 0.5 fmol),并且通过方差分析(p = 0.17)分析,靶向肽的 MMA 与丰度之间没有观察到系统趋势。第五个同位素的检测和线性度(即柱上 0.05 fmol)取决于肽和仪器扫描类型(MS1 与 PRM)。我们预计,该标准将成为一种强大的方法,用于进行仪器内性能监测/评估、技术开发和仪器间比较。