A Novel Method to Detect an Oxygenator Defect Prior to Cardiopulmonary Bypass Initiation.

作者信息

Fernandes Armindo, Laliberte Eric, Toledano Karine, Demers Philippe

机构信息

Perfusion Services, Montreal Heart Institute, Montréal, Canada.

Department of Anesthesiology, Hoôpital du Sacré-Cœur de Montréal, Montréal, Canada.

出版信息

J Extra Corpor Technol. 2015 Sep;47(3):180-2.

Abstract

Cardiopulmonary bypass (CPB) is a common practice in our era. The medical technology used for cardiac surgery goes through rigorous testing to ensure its safety. Unfortunately, it is not fail proof. Oxygenator failures are a rare occurrence but may lead to catastrophic events. We present a case where the preparation for initiating CPB was complicated by an oxygenator defect. After thorough examination, the oxygenator was found leaking from the gas exhaust port suggesting a disruption in continuity of the fibers. This was found by the vigilance of the perfusionist and a creative method to quickly assess the integrity of the oxygenation device. We describe a simple technique to help diagnose an oxygenator leak.

摘要

体外循环(CPB)在我们这个时代是一种常见的操作。用于心脏手术的医疗技术经过严格测试以确保其安全性。不幸的是,它并非万无一失。氧合器故障是罕见事件,但可能导致灾难性后果。我们报告一例因氧合器缺陷导致启动体外循环准备工作复杂化的病例。经过全面检查,发现氧合器从排气口泄漏,提示纤维的连续性中断。这是通过灌注师的警觉以及一种快速评估氧合装置完整性的创新方法发现的。我们描述一种有助于诊断氧合器泄漏的简单技术。

相似文献

6
Water condensation from gas outlet of oxygenator.氧合器气体出口处的水凝结。
J Card Surg. 2020 Aug;35(8):2039-2040. doi: 10.1111/jocs.14841. Epub 2020 Jul 11.
8
Testing of heat exchangers in membrane oxygenators using air pressure.
Perfusion. 2006 Mar;21(2):105-7. doi: 10.1191/0267659106pf856oa.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验