Submaximal cardiopulmonary thresholds on a robotics-assisted tilt table, a cycle and a treadmill: a comparative analysis.

作者信息

Saengsuwan Jittima, Nef Tobias, Laubacher Marco, Hunt Kenneth J

机构信息

Division of Mechanical Engineering, Department of Engineering and Information Technology, Institute for Rehabilitation and Performance Technology, Bern University of Applied Sciences, Pestalozzistrasse 20, 3400, Burgdorf, Switzerland.

ARTORG Center for Biomedical Engineering Research, Gerontechnology and Rehabilitation Research Group, University of Bern, Bern, Switzerland.

出版信息

Biomed Eng Online. 2015 Nov 10;14:104. doi: 10.1186/s12938-015-0099-0.

Abstract

BACKGROUND

The robotics-assisted tilt table (RATT), including actuators for tilting and cyclical leg movement, is used for rehabilitation of severely disabled neurological patients. Following further engineering development of the system, i.e. the addition of force sensors and visual bio-feedback, patients can actively participate in exercise testing and training on the device. Peak cardiopulmonary performance parameters were previously investigated, but it also important to compare submaximal parameters with standard devices. The aim of this study was to evaluate the feasibility of the RATT for estimation of submaximal exercise thresholds by comparison with a cycle ergometer and a treadmill.

METHODS

17 healthy subjects randomly performed six maximal individualized incremental exercise tests, with two tests on each of the three exercise modalities. The ventilatory anaerobic threshold (VAT) and respiratory compensation point (RCP) were determined from breath-by-breath data.

RESULTS

VAT and RCP on the RATT were lower than the cycle ergometer and the treadmill: oxygen uptake (V'O2) at VAT was [mean (SD)] 1.2 (0.3), 1.5 (0.4) and 1.6 (0.5) L/min, respectively (p < 0.001); V'O2 at RCP was 1.7 (0.4), 2.3 (0.8) and 2.6 (0.9) L/min, respectively (p = 0.001). High correlations for VAT and RCP were found between the RATT vs the cycle ergometer and RATT vs the treadmill (R on the range 0.69-0.80). VAT and RCP demonstrated excellent test-retest reliability for all three devices (ICC from 0.81 to 0.98). Mean differences between the test and retest values on each device were close to zero. The ventilatory equivalent for O2 at VAT for the RATT and cycle ergometer were similar and both were higher than the treadmill. The ventilatory equivalent for CO2 at RCP was similar for all devices. Ventilatory equivalent parameters demonstrated fair-to-excellent reliability and repeatability.

CONCLUSIONS

It is feasible to use the RATT for estimation of submaximal exercise thresholds: VAT and RCP on the RATT were lower than the cycle ergometer and the treadmill, but there were high correlations between the RATT vs the cycle ergometer and vs the treadmill. Repeatability and test-retest reliability of all submaximal threshold parameters from the RATT were comparable to those of standard devices.

摘要

背景

机器人辅助倾斜台(RATT),包括用于倾斜和周期性腿部运动的致动器,用于严重残疾神经科患者的康复治疗。随着该系统的进一步工程开发,即增加力传感器和视觉生物反馈,患者可以积极参与该设备上的运动测试和训练。此前已对心肺功能峰值参数进行了研究,但将次最大参数与标准设备进行比较也很重要。本研究的目的是通过与自行车测力计和跑步机比较,评估RATT用于估计次最大运动阈值的可行性。

方法

17名健康受试者随机进行6次最大个体化递增运动测试,在三种运动方式中的每种方式上各进行两次测试。通过逐次呼吸数据确定通气无氧阈值(VAT)和呼吸补偿点(RCP)。

结果

RATT上的VAT和RCP低于自行车测力计和跑步机:VAT时的摄氧量(V'O2)分别为[平均值(标准差)]1.2(0.3)、1.5(0.4)和1.6(0.5)L/分钟(p<0.001);RCP时的V'O2分别为1.7(0.4)、2.3(0.8)和2.6(0.9)L/分钟(p = 0.001)。在RATT与自行车测力计以及RATT与跑步机之间发现VAT和RCP具有高度相关性(相关系数R在0.69 - 0.80范围内)。对于所有三种设备,VAT和RCP都表现出极好的重测信度(组内相关系数ICC从0.81到0.98)。每种设备上测试值与重测值之间的平均差异接近零。RATT和自行车测力计在VAT时的氧通气当量相似,且均高于跑步机。所有设备在RCP时的二氧化碳通气当量相似。通气当量参数表现出中等至极好的信度和可重复性。

结论

使用RATT估计次最大运动阈值是可行的:RATT上的VAT和RCP低于自行车测力计和跑步机,但RATT与自行车测力计以及与跑步机之间存在高度相关性。RATT所有次最大阈值参数的可重复性和重测信度与标准设备相当。