一种便于右肾上腺静脉采血的同轴导丝-导管技术:76例患者的评估

A Coaxial Guide Wire-Catheter Technique to Facilitate Right Adrenal Vein Sampling: Evaluation in 76 Patients.

作者信息

Andrews James C, Thompson Scott M, Young William F

机构信息

Department of Radiology, Division of Vascular and Interventional Radiology, and the 200 First St. SW, Rochester, MN 55902.

Department of Radiology, Division of Vascular and Interventional Radiology, and the 200 First St. SW, Rochester, MN 55902.

出版信息

J Vasc Interv Radiol. 2015 Dec;26(12):1871-3. doi: 10.1016/j.jvir.2015.08.020.

Abstract

A technique to facilitate blood sampling from the right adrenal vein is described. Between May 2012 and February 2015, 148 adrenal vein sampling (AVS) procedures were attempted. In 72 procedures, a simple 5-F end-hole catheter was employed. In 76, a coaxial guide wire technique was used when blood could not be aspirated, whereby a 0.018-inch guide wire was passed through the catheter and into a branch of the right adrenal vein and the sample was drawn around the wire by using a side-arm adaptor. Successful sampling was achieved in 71 of the 72 catheter-only procedures (98.6%) and in 75 of the 76 coaxial wire-assisted procedures (98.7%). This simple technique may eliminate the need for multiple catheter exchanges during AVS.

摘要

本文描述了一种便于从右肾上腺静脉采血的技术。在2012年5月至2015年2月期间,共尝试了148例肾上腺静脉采样(AVS)操作。其中72例操作使用了简单的5F端孔导管。另外76例操作在无法抽吸到血液时采用了同轴导丝技术,即通过导管插入一根0.018英寸的导丝进入右肾上腺静脉的一个分支,然后使用侧臂适配器围绕导丝抽取样本。仅使用导管的72例操作中有71例成功采样(98.6%),同轴导丝辅助的76例操作中有75例成功采样(98.7%)。这种简单的技术可能无需在AVS期间多次更换导管。