Jennings Jason M, Randell Timmothy R, Green Cynthia L, Zheng Guoyan, Wellman Samuel S
Colorado Joint Replacement, Denver, Colorado.
Department of Orthopaedic Surgery, Duke University Medical Center, Durham, North Carolina.
J Arthroplasty. 2016 Mar;31(3):658-61. doi: 10.1016/j.arth.2015.09.052. Epub 2015 Oct 9.
Acetabular cup malpositioning during total hip arthroplasty may lead to impingement, instability, wear-induced osteolysis, and increased rates of revision surgery. The purpose of this study was to independently evaluate the accuracy of acetabular cup orientation using a novel mechanical navigation device.
An adjustable, reusable mechanical navigation device was used in a consecutive series of patients. Angles (inclination and anteversion) were measured by 2 independent reviewers with a validated 2-dimensional/3-dimensional matching application using a preoperative computed tomographic scan and a postoperative plain film.
There were no outliers for inclination or anteversion for errors within 10° of the preoperative plan. There were 6 (12.8%) outliers for inclination and 11 (23.4%) for anteversion for errors within 5° of the preoperative plan. All patients simultaneously met both targets when an outlier was considered 10°, whereas 30 (63.8%) simultaneously met both targets when the criteria was tightened to 5°. The absolute errors for both inclination and anteversion were significantly less than both 5° and 10° (P<.001).
This device reliably navigates acetabular cup inclination and anteversion individualized for each patient to within 10° of a preoperative plan. Outliers increased when this criteria is tightened to 5° but still appears to be more accurate than conventional acetabular cup component placement. Further research is warranted to assess the clinical impact of reducing outliers using this device.
全髋关节置换术中髋臼杯位置不当可能导致撞击、不稳定、磨损性骨溶解以及翻修手术率增加。本研究的目的是使用一种新型机械导航设备独立评估髋臼杯方向的准确性。
在一系列连续的患者中使用了一种可调节、可重复使用的机械导航设备。由两名独立的评估者使用经过验证的二维/三维匹配应用程序,通过术前计算机断层扫描和术后平片测量角度(倾斜度和前倾角)。
术前计划误差在10°以内时,倾斜度或前倾角均无异常值。术前计划误差在5°以内时,倾斜度有6个(12.8%)异常值,前倾角有11个(23.4%)异常值。当将异常值视为10°时,所有患者同时达到两个目标,而当标准收紧至5°时,30名(63.8%)患者同时达到两个目标。倾斜度和前倾角的绝对误差均显著小于5°和10°(P<0.001)。
该设备可将每个患者个体化的髋臼杯倾斜度和前倾角可靠地导航至术前计划的10°范围内。当标准收紧至5°时,异常值增加,但似乎仍比传统髋臼杯组件放置更准确。有必要进行进一步研究以评估使用该设备减少异常值的临床影响。