Yonezawa Y, Caldwell W M, Schadt J C, Hahn A W
Biomed Sci Instrum. 1989;25:107-11.
An ultrasonic blood flowmeter telemetry system, using only two integrated circuits and one transistor, has been designed for chronic unanesthetized animal studies. Standard components and construction techniques are utilized. The CW Doppler system employs LZT piezoelectric 10 MHZ transmitting and receiving crystals held in a lightweight polystyrene arterial transducer. The audio frequency, FM Doppler signal modulates an FM oscillator-transmitter to produce an FM/FM radio frequency carrier which is transmitted to an FM receiver at a remote location. The transmitted Doppler audio flow signal is demodulated at the receiver by a zero crossing detector. Power consumption is 48 mw. The 12.4 cc package volume can be further reduced for animal implantation.
一种仅使用两个集成电路和一个晶体管的超声血流仪遥测系统已被设计用于慢性未麻醉动物研究。采用了标准组件和构建技术。连续波多普勒系统采用LZT压电10兆赫发射和接收晶体,置于轻质聚苯乙烯动脉换能器中。音频调频多普勒信号调制一个调频振荡器 - 发射机,以产生一个调频/调频射频载波,该载波被传输到远程位置的调频接收机。发射的多普勒音频血流信号在接收机处由过零检测器解调。功耗为48毫瓦。12.4立方厘米的封装体积可进一步减小以用于动物植入。