Heery Brendan, Briciu-Burghina Ciprian, Zhang Dian, Duffy Gillian, Brabazon Dermot, O'Connor Noel, Regan Fiona
DCU Water Institute, Marine and Environmental Sensing Technology Hub (MESTECH), NCSR, Dublin City University (DCU), Glasnevin, Dublin 9, Ireland.
School of Electronic Engineering, Dublin City University, Ireland.
Talanta. 2016;148:75-83. doi: 10.1016/j.talanta.2015.10.035. Epub 2015 Oct 21.
A sensitive field-portable fluorimeter with incubating capability and triplicate sample chambers was designed and built. The system was optimised for the on-site analysis of E. coli in recreational waters using fluorescent based enzyme assays. The target analyte was β-D-Glucuronidase (GUS) which hydrolyses a synthetic substrate 6-Chloro-4-Methyl-Umbelliferyl-β-D-Glucuronide (6-CMUG) to release the fluorescent molecule 6-Chloro-4-Methyl-Umbelliferyl (6-CMU). The system was calibrated with 6-CMU standards. A LOD of 5 nM and a resolution of less than 1 nM was determined while enzyme kinetic tests showed detection of activities below 1 pmol min(-1) mL(-1) of sample. A field portable sample preparation, enzyme extraction protocol and continuous assay were applied with the system to analyse freshwater and marine samples. Results from a one day field trial are shown which demonstrated the ability of the system to deliver results on-site within a 75 min period.
设计并制造了一种具有孵育功能和一式三份样品室的灵敏的现场便携式荧光计。该系统针对娱乐用水中大肠杆菌的现场分析进行了优化,采用基于荧光的酶分析法。目标分析物是β-D-葡萄糖醛酸酶(GUS),它水解合成底物6-氯-4-甲基伞形酮基-β-D-葡萄糖醛酸苷(6-CMUG)以释放荧光分子6-氯-4-甲基伞形酮(6-CMU)。该系统用6-CMU标准品进行校准。测定的检测限为5 nM,分辨率小于1 nM,而酶动力学测试表明能检测到样品中低于1 pmol min(-1) mL(-1)的活性。该系统应用了现场便携式样品制备、酶提取方案和连续测定法来分析淡水和海洋样品。展示了一天现场试验的结果,证明该系统有能力在75分钟内现场给出结果。