To evaluate quantification of early fatty infiltration in supraspinatus muscles with magnetic resonance (MR) imaging using a T2*-corrected multi-echo 3D-gradient-echo Dixon-based sequence (multi-echo Dixon) and compare it to proton-MR-spectroscopy.
Sixty subjects (mean age 46 years, 41 men) with good supraspinatus muscle quality on 1.5 T MR imaging were included. Fat percentage (FP) in the supraspinatus muscle was quantified using a multi-echo Dixon compared to single-voxel MR spectroscopy as reference standard. In 18 subjects the multi-echo Dixon was repeated to assess test-retest reliability. Measurements based on multi-echo Dixon were performed by two independent readers by placing regions-of-interest (ROIs) in the supraspinatus muscle corresponding to the MR-spectroscopy voxel. Intraclass and concordance correlation coefficients (ICC/CCC) were used for statistical analysis.
Test-retest reliability was substantial for reader 1 (ICC = 0.757) and almost perfect for reader 2 (ICC = 0.873). Inter-reader reliability for multi-echo Dixon was almost perfect (ICC = 0.893, P < .0005). Mean FP in all 60 subjects with multi-echo Dixon was 3.5 ± 1.6 for reader 1, 3.7 ± 1.8 for reader 2, and 2.8 ± 1.4 with MR spectroscopy. Correlation between multi-echo Dixon and MR spectroscopy was moderate (CCC = 0.641).
The multi-echo Dixon sequence is a reliable method and comparable to MR-spectroscopy for quantification of low levels of fatty infiltration in the supraspinatus muscle.
• Multi-echo Dixon for low fat quantification in muscles is reliable. • Multi-echo Dixon low fat quantification is comparable to single-voxel MR spectroscopy • Multi-echo Dixon detects substantial differences in fatty infiltration within Goutallier 0-1.
使用基于T2*校正的多回波3D梯度回波狄克逊序列(多回波狄克逊)通过磁共振(MR)成像评估冈上肌早期脂肪浸润的量化,并将其与质子MR波谱进行比较。
纳入60名在1.5T MR成像上冈上肌质量良好的受试者(平均年龄46岁,41名男性)。以单像素MR波谱作为参考标准,使用多回波狄克逊量化冈上肌中的脂肪百分比(FP)。对18名受试者重复多回波狄克逊以评估重测信度。基于多回波狄克逊的测量由两名独立的阅片者通过在冈上肌中对应于MR波谱体素的位置放置感兴趣区域(ROI)来进行。组内相关系数和一致性相关系数(ICC/CCC)用于统计分析。
阅片者1的重测信度较高(ICC = 0.757),阅片者2的重测信度几乎完美(ICC = 0.873)。多回波狄克逊的阅片者间信度几乎完美(ICC = 0.893,P <.0005)。在所有60名受试者中,阅片者1使用多回波狄克逊测得的平均FP为3.5±1.6,阅片者2为3.7±1.8,MR波谱测得的为2.8±1.4。多回波狄克逊与MR波谱之间的相关性为中等(CCC = 0.641)。
多回波狄克逊序列是一种可靠的方法,在量化冈上肌低水平脂肪浸润方面与MR波谱相当。
• 多回波狄克逊用于肌肉中低脂肪量化是可靠的。• 多回波狄克逊低脂肪量化与单像素MR波谱相当。• 多回波狄克逊可检测出Goutallier 0 - 1期内脂肪浸润的显著差异。