Massaroni C, Schena E, Saccomandi P, Morrone M, Sterzi S, Silvestri S
Annu Int Conf IEEE Eng Med Biol Soc. 2015 Aug;2015:1291-4. doi: 10.1109/EMBC.2015.7318604.
Opto-electronic Plethysmography (OEP) is a motion analysis system used to measure chest wall kinematics and to indirectly evaluate respiratory volumes during breathing. Its working principle is based on the computation of marker displacements placed on the chest wall. This work aims at evaluating the accuracy and precision of OEP in measuring displacement in the range of human chest wall displacement during quiet breathing. OEP performances were investigated by the use of a fully programmable chest wall simulator (CWS). CWS was programmed to move 10 times its eight shafts in the range of physiological displacement (i.e., between 1 mm and 8 mm) at three different frequencies (i.e., 0.17 Hz, 0.25 Hz, 0.33 Hz). Experiments were performed with the aim to: (i) evaluate OEP accuracy and precision error in recording displacement in the overall calibrated volume and in three sub-volumes, (ii) evaluate the OEP volume measurement accuracy due to the measurement accuracy of linear displacements. OEP showed an accuracy better than 0.08 mm in all trials, considering the whole 2m(3) calibrated volume. The mean measurement discrepancy was 0.017 mm. The precision error, expressed as the ratio between measurement uncertainty and the recorded displacement by OEP, was always lower than 0.55%. Volume overestimation due to OEP linear measurement accuracy was always <; 12 mL (<; 3.2% of total volume), considering all settings.
光电体积描记法(OEP)是一种运动分析系统,用于测量胸壁运动学并间接评估呼吸过程中的呼吸量。其工作原理基于对放置在胸壁上的标记位移的计算。这项工作旨在评估OEP在测量安静呼吸时人体胸壁位移范围内的位移时的准确性和精确性。通过使用完全可编程的胸壁模拟器(CWS)来研究OEP的性能。CWS被编程为在生理位移范围内(即1毫米至8毫米之间)以三种不同频率(即0.17赫兹、0.25赫兹、0.33赫兹)移动其八个轴10次。进行实验的目的是:(i)评估OEP在记录整个校准体积和三个子体积中的位移时的准确性和精确性误差,(ii)评估由于线性位移的测量精度导致的OEP体积测量准确性。考虑整个2立方米的校准体积,在所有试验中OEP的准确性均优于0.08毫米。平均测量差异为0.017毫米。以测量不确定度与OEP记录的位移之比表示的精确性误差始终低于0.55%。考虑所有设置,由于OEP线性测量精度导致的体积高估始终<12毫升(<总体积的3.2%)。