Evans D H, Schlindwein F S, Levene M I
Department of Medical Physics and Clinical Engineering, Leicester Royal Infirmary, UK.
Ultrasound Med Biol. 1989;15(5):429-35. doi: 10.1016/0301-5629(89)90095-1.
It is usual practice to calculate the mean velocity of blood flow in neonatal cerebral vessels from the intensity weighted mean (IWM) Doppler shift signal. Theoretically however the maximum frequency envelope could be used for similar purposes, and indeed may have certain advantages. The purpose of this study was to confirm the suitability of the maximum frequency method. Good quality Doppler recordings from the anterior cerebral and middle cerebral arteries of both term and very low birth weight babies were analyzed by both methods and compared. A small number of recordings were then deliberately degraded, either by the addition of noise or by the use of inappropriate filters, and reanalyzed. The results of these comparisons suggest that the maximum frequency follower should be the follower of choice for neonatal cerebral blood flow studies.
从强度加权平均(IWM)多普勒频移信号计算新生儿脑血管中血流平均速度是惯例。然而从理论上讲,最大频率包络可用于类似目的,而且确实可能具有某些优势。本研究的目的是证实最大频率法的适用性。通过这两种方法对足月儿和极低出生体重儿的大脑前动脉和大脑中动脉的高质量多普勒记录进行了分析并比较。然后,通过添加噪声或使用不合适的滤波器故意降低少数记录的质量,并重新进行分析。这些比较结果表明,最大频率跟踪器应是新生儿脑血流研究的首选跟踪器。