Rivett Cat, Dixon Matthew, Matthews Lucy, Rowles Nick
Clinical and Radiation Physics, Plymouth Hospitals NHS Trust, Level 2, Derriford Hospital, Derriford Road, Plymouth, Devon PL6 8DH, UK
Clinical and Radiation Physics, Plymouth Hospitals NHS Trust, Level 2, Derriford Hospital, Derriford Road, Plymouth, Devon PL6 8DH, UK.
Radiat Prot Dosimetry. 2016 Dec;172(4):443-452. doi: 10.1093/rpd/ncv540. Epub 2016 Jan 13.
In light of the proposal from the International Commission on Radiological Protection for a lowered eye dose limit, now adopted by a European Union Council Directive, lead glasses may be required for some staff in interventional radiology to ensure that occupational exposure is as low as reasonably practicable. To investigate the lens protection offered from various models of lead glasses exposed to X-rays coming from a source to the left and below, calibrated radiochromic film was positioned in the lens area of a head phantom. When the source-to-eye angles were large, the dose reduction factors (the ratio of eye dose without protection to dose with protection) to the right lens area were much lower than to the left lens area, particularly with smaller-lensed glasses, due to gaps in protection between the face and the glasses. The results of this study reiterate the importance of employers providing eyewear based on the morphology of, and fit to, individual workers' faces.