Field John R, Callary Stuart A, Solomon Lucian B, Stamenkov Roumen, McGee Margaret A, Howie Donald W
John R. Field, Centre for Orthopaedic and Trauma Research, Level 4, Bice Building, Royal Adelaide Hospital, Adelaide, South Australia, Australia, Phone: +61 422 469 263, E-mail:
Vet Comp Orthop Traumatol. 2016;29(2):125-30. doi: 10.3415/VCOT-15-04-0063. Epub 2016 Jan 20.
Hemiarthroplasty induces degenerative changes in the hip joint, which are difficult to evaluate in vivo. Radiostereometric analysis (RSA) is a radiographic measurement technique that has recently been used to measure acetabular cartilage wear in vivo. The aim of the study was to measure acetabular cartilage wear, using this technique, in an ovine model during the first 14 weeks post-implantation.
Measurements of three-dimensional femoral head migration, combined with visual assessments at necropsy and safranin O staining for cartilage integrity, were undertaken.
Mean femoral head migration during the first six weeks was 0.525 mm in the medial, 0.144 mm in the cranial, and 0.517 mm in the dorsal direction. The majority of this migration was confirmed to be cartilage wear in the medial and dorsal aspects of the acetabulum at necropsy and with subsequent histological evaluation depicting significant cartilage degeneration.
Radiostereometric analysis is the current gold standard technique for in vivo assessment of implant migration following total hip replacement. This study has utilized RSA to quantify the amount of early cartilage wear in vivo, which was supported by ex vivo evaluations. Accurately measuring the amount of cartilage wear will allow future studies to compare component material and design characteristics prior to clinical use.
半髋关节置换术会引发髋关节的退行性变化,而这种变化在活体中难以评估。放射立体测量分析(RSA)是一种放射成像测量技术,最近已被用于在活体中测量髋臼软骨磨损情况。本研究的目的是运用该技术,在绵羊模型植入后的前14周内测量髋臼软骨磨损情况。
进行了三维股骨头移位测量,并结合尸检时的视觉评估以及用于软骨完整性评估的番红O染色。
前六周内股骨头的平均移位情况为:内侧0.525毫米、头侧0.144毫米、背侧0.517毫米。尸检及后续组织学评估显示髋臼内侧和背侧的大部分移位被确认为软骨磨损,且存在显著的软骨退变。
放射立体测量分析是目前全髋关节置换术后植入物移位活体评估的金标准技术。本研究利用RSA对活体中早期软骨磨损量进行了量化,这得到了体外评估的支持。准确测量软骨磨损量将使未来的研究能够在临床使用前比较假体材料和设计特征。