Tharmviboonsri T, Riansuwan K, Nitising A, Mahaisavariya B
Department of Orthopaedic Surgery and Rehabilitation, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
Department of Surgery, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
Eur J Trauma Emerg Surg. 2012 Jun;38(3):307-11. doi: 10.1007/s00068-011-0165-7. Epub 2011 Nov 19.
To measure the radiation exposure at the knee and surrounding area during the use of 3D fluoroscopy.
An experimental study was conducted by using a human cadaveric knee as a focus point for the 3D fluoroscope. An isocentric C-arm fluoroscope machine was applied on the lateral side of the knee. The radiation dosage at the focus point and surrounding area was measured. The mean radiation exposure in each location was compared between low- and high-resolution scanning.
The mean radiation sustained at the focus point was 44.0 ± 5.6 μSv and 20.0 ± 1.0 μSv in high- and low-resolution scanning, respectively. Radiation exposure on the opposite side of the C-arm machine was found to be higher than that on the other locations with the same distance from the focus point. In low-resolution scanning, radiation could not be detected beyond 75 cm from the focus point at the proximal, distal and same side of the machine. Radiation could be measured at a distance of up to 1.25 m on the opposite side of the machine. In high-resolution scanning, radiation could be measured at a distance of up to 1 m at the proximal, distal and same side of the C-arm, but up to 1.5 m on the opposite side.
Radiation exposure during 3D fluoroscopy of the knee decreases with increasing distance from the focus point. A higher number of scans in the high-resolution mode causes greater radiation exposure. In isocentric 3D fluoroscopy of the knee, a safe zone is located at least 1.5 m away from the focus point.
测量使用三维荧光透视时膝关节及周围区域的辐射暴露量。
以人体尸体膝关节作为三维荧光透视的聚焦点进行实验研究。在膝关节外侧使用等中心C型臂荧光透视机。测量聚焦点及周围区域的辐射剂量。比较低分辨率和高分辨率扫描时各位置的平均辐射暴露量。
高分辨率和低分辨率扫描时,聚焦点处的平均辐射量分别为44.0±5.6微希沃特和20.0±1.0微希沃特。发现C型臂机器对面的辐射暴露高于距聚焦点相同距离的其他位置。在低分辨率扫描中,在机器近端、远端和同侧距聚焦点75厘米以外无法检测到辐射。在机器对面距离达1.25米处可测量到辐射。在高分辨率扫描中,在C型臂近端、远端和同侧距离达1米处可测量到辐射,但在对面距离达1.5米处可测量到辐射。
膝关节三维荧光透视期间的辐射暴露随距聚焦点距离的增加而降低。高分辨率模式下扫描次数越多,辐射暴露越大。在膝关节等中心三维荧光透视中,安全区位于距聚焦点至少1.5米处。