Ganser Antonia, Thompson Rosemary E, Tami Ivan, Neuhoff Dirk, Steiner Adrian, Ito Keita
AO Research Institute, Davos Platz, Switzerland.
Faculty of Veterinary Medicine, University of Berne, Berne, Switzerland.
Eur J Trauma Emerg Surg. 2007 Feb;33(1):59-68. doi: 10.1007/s00068-007-6053-5. Epub 2007 Feb 27.
To compare the clinical benefits of stainless steel (SS) to titanium (Ti) on reducing pin track irritation/infection and pin loosening during external fracture fixation.
A tibial gap osteotomy was created in 17 sheep and stabilized with four Schanz screws of either SS or Ti and an external fixation frame. Over the 12 week observation period, pin loosening was assessed by grading the radiolucency around the pins and measuring the extraction torque on pin removal at sacrifice. Irritation/infection was assessed with weekly clinical pin track grading. A histological analysis of the tissue adjacent to the pin site was made to assess biocompatibility.
A statistically non-significant trend for less bone resorption around Ti pins was found during the early observation period. However, at sacrifice, there was no difference between the two materials. Also, there was no difference in the extraction torque, and there was similar remodeling and apposition of the bone around the pins. A statistically non-significant trend for more infection about SS pins at sacrifice was found. Histology showed a slightly higher prevalence of reactionary cells in SS samples, but was otherwise not much different than around Ti pins.
There is no clinically relevant substantial advantage in using either SS or Ti pins on reducing pin loosening or pin track irritation/infection.
比较不锈钢(SS)与钛(Ti)在减少外固定治疗骨折期间针道刺激/感染及钢针松动方面的临床效果。
对17只绵羊行胫骨截骨术,用4枚SS或Ti斯氏针及外固定架固定。在12周的观察期内,通过对钢针周围透亮区分级及在处死时测量拔除钢针的拔出扭矩来评估钢针松动情况。每周对针道进行临床分级以评估刺激/感染情况。对针道周围组织进行组织学分析以评估生物相容性。
在观察早期,发现Ti针周围骨吸收较少,但差异无统计学意义。然而,处死时,两种材料之间无差异。此外,拔出扭矩无差异,针周围骨的重塑和贴附相似。处死时发现SS针感染略多,但差异无统计学意义。组织学显示SS样本中反应性细胞的发生率略高,但与Ti针周围情况无太大差异。
在减少钢针松动或针道刺激/感染方面,使用SS或Ti针在临床上并无显著优势。