Suppr超能文献

心肌灌注成像检测多支冠状动脉疾病的准确性:一项心脏CZT研究。

Accuracy of myocardial perfusion imaging in detecting multivessel coronary artery disease: A cardiac CZT study.

作者信息

Gimelli Alessia, Liga Riccardo, Duce Valerio, Kusch Annette, Clemente Alberto, Marzullo Paolo

机构信息

Fondazione Toscana Gabriele Monasterio, Via Moruzzi, 1, 56124, Pisa, Italy.

Cardio-thoracic and Vascular Department, University Hospital of Pisa, Pisa, Italy.

出版信息

J Nucl Cardiol. 2017 Apr;24(2):687-695. doi: 10.1007/s12350-015-0360-8. Epub 2016 Feb 3.

Abstract

BACKGROUND

Myocardial perfusion imaging (MPI) performed on traditional single-photon emission computed-tomography cameras has been shown to have a sub-optimal accuracy in detecting multivessel coronary artery disease (CAD).

METHODS

Six-hundred and ninety-five patients were submitted to MPI on a novel cadmium-zinc-telluride (CZT) camera and coronary angiography. A coronary stenosis >70% was considered obstructive. In every patient, the summed stress score (SSS) was computed. Moreover, the regional stress scores were also calculated for every coronary territory.

RESULTS

Four-hundred and forty-one patients had obstructive CAD in one (28%), two (19%), or three (17%) vessels. At per-patient analysis, the SSS showed a significant accuracy in detecting obstructive CAD (AUC 0.87, P < .001). Specifically, its accuracy was maintained also in patients with double (AUC 0.83; P < .001) or triple-vessels disease (AUC 0.79, P < .001), where CZT was able to correctly identify CAD extent in 64% of patients. On a per-vessel basis, CZT confirmed its high accuracy in detecting obstructive CAD (AUC 0.88, P < .001), independently from the involved coronary vessel.

CONCLUSIONS

MPI performed on a CZT camera is highly accurate in detecting obstructive CAD, independently from the coronary artery involved and the overall disease burden.

摘要

背景

在传统单光子发射计算机断层扫描相机上进行的心肌灌注成像(MPI)已被证明在检测多支冠状动脉疾病(CAD)时准确性欠佳。

方法

695例患者在新型碲化镉锌(CZT)相机上进行了MPI检查及冠状动脉造影。冠状动脉狭窄>70%被视为阻塞性病变。为每位患者计算了总应力评分(SSS)。此外,还为每个冠状动脉区域计算了局部应力评分。

结果

441例患者存在单支(28%)、双支(19%)或三支(17%)血管的阻塞性CAD。在患者个体分析中,SSS在检测阻塞性CAD方面显示出显著的准确性(曲线下面积[AUC]为0.87,P <.001)。具体而言,在双支血管疾病(AUC 0.83;P <.001)或三支血管疾病(AUC 0.79,P <.001)患者中其准确性也得以保持,其中CZT能够在64%的患者中正确识别CAD范围。在每支血管的基础上,CZT证实其在检测阻塞性CAD方面具有很高的准确性(AUC 0.88,P <.001),与受累冠状动脉血管无关。

结论

在CZT相机上进行的MPI在检测阻塞性CAD方面具有很高的准确性,与受累冠状动脉及整体疾病负担无关。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验