Suppr超能文献

使用心肌灌注 SPECT 数据获得的虚拟平面图像进行肺心比值分析:一项体模和临床研究。

Lung-to-heart ratio analysis using virtual planar images obtained from myocardial perfusion SPECT data: A phantom and clinical studies.

机构信息

Department of Radiology, National Cerebral and Cardiovascular Center, Suita, Osaka, Japan.

出版信息

J Nucl Cardiol. 2023 Oct;30(5):1959-1967. doi: 10.1007/s12350-023-03233-3. Epub 2023 Mar 14.

Abstract

BACKGROUNDS

The lung-to-heart ratio (L/H ratio) in myocardial perfusion scintigraphy (MPS) is a useful marker that complements the sensitivity of ischemia detection. However, it requires planar imaging acquired following a separate protocol in addition to single-photon emission computed tomography (SPECT). We developed a novel method for constructing virtual planar image (VPI) from SPECT data.

METHODS

Myocardial phantoms using Tl-201 were built with different amounts of radioactivity in the lungs. SPECT data and conventional planar images of these phantoms were collected with an Anger-type gamma camera. VPIs were constructed by adding all coronal images reconstructed from SPECT data. The clinical utility of VPIs obtained from 52 patients who underwent MPS with Tc-99m sestamibi was evaluated.

RESULTS

The radioactivity linearity of VPIs was satisfactory, with a correlation coefficient of r ≥ .99 between the measured amounts of radioactivity and image counts. The L/H ratios obtained from VPI analysis were strongly correlated with those of conventional planar images with a correlation coefficient of r ≥ .99 in the phantom study and r = .929 in clinical application.

CONCLUSION

The accuracy of VPI-based L/H ratio analysis was comparable to that of conventional planar image-based analysis. VPIs could be used as an alternative method of obtaining planar images in clinical settings.

摘要

背景

心肌灌注闪烁显像(MPS)中的肺心比(L/H 比)是一种有用的标志物,可补充缺血检测的敏感性。然而,它需要在进行单光子发射计算机断层扫描(SPECT)之外,采用单独的协议获取平面成像。我们开发了一种从 SPECT 数据构建虚拟平面图像(VPI)的新方法。

方法

使用 Tl-201 构建具有不同肺部放射性活度的心肌体模。使用 Anger 型伽玛相机收集这些体模的 SPECT 数据和常规平面图像。通过添加从 SPECT 数据重建的所有冠状图像来构建 VPI。评估了从 52 名接受 Tc-99m sestamibi MPS 的患者获得的 VPI 的临床实用性。

结果

VPI 的放射性线性度令人满意,放射性活度的测量值与图像计数之间的相关系数 r≥.99。在体模研究中,VPI 分析得出的 L/H 比值与传统平面图像分析得出的 L/H 比值具有很强的相关性,相关系数 r≥.99,在临床应用中 r=0.929。

结论

基于 VPI 的 L/H 比分析的准确性与基于传统平面图像的分析相当。VPI 可作为临床环境中获得平面图像的替代方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验