Mizuno Shiho, Sonoda Shigeru, Takeda Kotaro, Maeshima Shinichiro
School of Medicine, Department of Rehabilitation Medicine II, Fujita Health University, Mie, Japan.
School of Medicine, Department of Rehabilitation Medicine II, Fujita Health University, Mie, Japan.
J Stroke Cerebrovasc Dis. 2016 Apr;25(4):946-53. doi: 10.1016/j.jstrokecerebrovasdis.2015.12.036. Epub 2016 Feb 3.
Quantification of increased muscle tone for patients with spasticity has been performed to date using various devices to replace the manual scales, such as the modified Ashworth scale or the Tardieu scale. We developed a device that could measure resistive plantar flexion (PF) torque of the ankle during passive dorsiflexion (DF) as an indicator of muscle tone of ankle plantar flexors.
The primary objective was to explore the test-retest intrarater reliability of a custom-built device. Participants were 11 healthy subjects (7 men, 4 women; mean age 47.0 years) and 22 patients with poststroke hemiplegia (11 hemorrhagic, 11 ischemic; 14 men, 8 women; mean age 57.2 years). The device was affixed to the ankle. Subjects were seated with knees either flexed or extended. The ankle was passively dorsiflexed from 20° of PF to more than 10° of DF at 5°/second (slow stretch) or 90°/second (fast stretch). Angle and torque were measured twice during the stretches. The intraclass correlation coefficients (ICCs) of torque at 10° of DF (T10) in the 4 conditions-slow and fast stretches with knee flexed or extended-were calculated.
The T10 ICCs of the 4 conditions were .95-.99 in both groups. The healthy subjects showed significantly higher T10 of knee extension than of knee flexion during slow and fast stretches. The patients showed increased velocity-dependent torque during fast stretches.
Excellent reliability was observed. The device is suitable for measuring resistive PF torque during passive stretch in a flexed knee condition.
迄今为止,已使用各种设备(如改良的Ashworth量表或Tardieu量表)替代手动量表来量化痉挛患者增加的肌张力。我们开发了一种设备,该设备可以测量被动背屈(DF)过程中踝关节的抵抗性跖屈(PF)扭矩,作为踝关节跖屈肌肌张力的指标。
主要目的是探索一种定制设备的重测信度。参与者包括11名健康受试者(7名男性,4名女性;平均年龄47.0岁)和22名中风后偏瘫患者(11名出血性,11名缺血性;14名男性,8名女性;平均年龄57.2岁)。该设备固定在踝关节上。受试者坐着,膝盖可弯曲或伸直。踝关节以5°/秒(缓慢拉伸)或90°/秒(快速拉伸)的速度从PF的20°被动背屈至超过DF的10°。在拉伸过程中对角度和扭矩进行两次测量。计算了在DF为10°(T10)时,在4种条件下(膝盖弯曲或伸直的缓慢和快速拉伸)扭矩的组内相关系数(ICC)。
两组在4种条件下的T10 ICC均为0.95 - 0.99。健康受试者在缓慢和快速拉伸过程中,膝盖伸直时的T10明显高于膝盖弯曲时。患者在快速拉伸过程中表现出速度依赖性扭矩增加。
观察到极好的可靠性。该设备适用于在屈膝状态下被动拉伸过程中测量抵抗性PF扭矩。