Suppr超能文献

扫频源数字全息相位显微镜。

Swept source digital holographic phase microscopy.

作者信息

Chen Shichao, Ryu Junghyun, Lee Kiho, Zhu Yizheng

出版信息

Opt Lett. 2016 Feb 15;41(4):665-8. doi: 10.1364/OL.41.000665.

Abstract

We propose a swept source-based digital holographic phase microscopy technique. By scanning source wavelength, a series of on-axis interferograms can be obtained for accurate determination of the sample phase using spectral domain interferometry. With these sample spectra, sources of undesirable interference artifacts, often significant in holographic systems, can be identified and avoided by placing the sample signal at a spectral frequency with a clean background. Pathlength sensitivity better than 0.3 nm can, thus, be achieved. The quantitative pathlength image of live sperm cells is obtained with clear identification of morphological features. In addition, the availability of sample spectrum also permits the retrieval of its spectroscopic information. The wavelength-dependent refractive indices of indocyanine green solution are obtained to demonstrate this capability.

摘要

我们提出了一种基于扫频光源的数字全息相位显微镜技术。通过扫描光源波长,可以获得一系列同轴干涉图,以便使用光谱域干涉测量法精确测定样品相位。利用这些样品光谱,通过将样品信号置于具有干净背景的光谱频率处,可以识别并避免在全息系统中通常很显著的不良干涉伪影源。因此,可以实现优于0.3 nm的光程灵敏度。获得了活精子细胞的定量光程图像,并清晰地识别了其形态特征。此外,样品光谱的可用性还允许检索其光谱信息。通过获得吲哚菁绿溶液的波长相关折射率来证明这种能力。