COMPARISON STUDY OF VARIOUS PLASTICS AS THE WALL MATERIAL OF THGEM-BASED MICRODOSEMETERS FOR FAST NEUTRON MEASUREMENTS.

作者信息

Moslehi A, Raisali G, Lamehi M

机构信息

Radiation Applications Research School, Nuclear Science and Technology Research Institute, AEOI, PO Box 11365-3486, Tehran, Iran.

Physics and Accelerators Research School, Nuclear Science and Technology Research Institute, AEOI, PO Box 11365-3486, Tehran, Iran.

出版信息

Radiat Prot Dosimetry. 2017 Apr 15;173(4):286-292. doi: 10.1093/rpd/ncw008.

Abstract

To find appropriate substitutions for the expensive plastics of A-150 and rexolite used in the construction of thick gas electron multiplier (THGEM)-based tissue-equivalent proportional counters, in the present work, the responses of a THGEM-based microdosimetric detector made of A-150 and rexolite and three others composed of plexiglas (PMMA), polyethylene and polystyrene plastics as the wall materials have been compared. Lineal energy distribution, frequency-averaged lineal energy, dose-averaged lineal energy, mean quality factor and dose-equivalent for 0.1, 1 and 10 MeV neutrons and also for 241Am-Be neutrons are calculated using Geant4 simulation toolkit. Frequency-averaged lineal energy, dose-averaged lineal energy, mean quality factor and dose-equivalent values for all plastics are found similar. In addition, the response of an indigenously constructed microdosemeter with PMMA walls is also measured for 241Am-Be neutrons. The experimental results are in good agreement with the simulation predictions. Conclusively, it was found that the three considered plastics can be used as good candidates instead of A-150 and rexolite plastics in fast neutron microdosimetry.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验