Dual chamber pacing mode in an atrial antitachycardia pacing device without a ventricular lead - A necessary evil.

作者信息

Noheria Amit, Friedman Paul A, Asirvatham Samuel J, McLeod Christopher J

机构信息

Division of Cardiovascular Diseases, Mayo Clinic, Rochester, MN, USA.

出版信息

Indian Pacing Electrophysiol J. 2015 Jul 29;15(2):133-7. doi: 10.1016/j.ipej.2015.07.012. eCollection 2015 Mar-Apr.

Abstract

We present a case of a single chamber atrial pacemaker implanted for sinus node dysfunction and treatment of macroreentrant atrial tachycardias with atrial antitachycardia pacing. The patient presented with sustained atrial tachycardia above the detection rate, however, the device was unable to detect the tachycardia and did not deliver the programmed therapy. We discuss the nuances of the atrial tachyarrhythmia detection algorithms, and the programming strategies to maximize detection of atrial arrhythmias in a single chamber atrial pacemaker.