Shortland A P, Cochrane T
Department of Medical Physics and Clinical Engineering, Royal Hallamshire Hospital, Sheffield, United Kingdom.
Ultrasound Med Biol. 1989;15(8):737-48. doi: 10.1016/0301-5629(89)90114-2.
This paper describes a microcomputer controlled pump which generates pulsatile flows similar to those found in the human peripheral circulation. Continuous wave Doppler ultrasound was used to investigate the flows generated by the pump and the behaviour of diagnostic indices derived from the spectra was examined. Sonograms were recorded from elastic and rigid tubes with various degrees of axisymmetric constriction. Heart rate, stroke volume, severity of vessel constriction, vessel wall elasticity, distal resistance, and systemic peripheral impedance were varied in turn and the resulting Doppler spectra compared. Indices considered were the pulsatility index, spectral broadening index and peck Doppler frequency. In general, the indices lacked the sensitivity to detect low to moderate levels of disease.
本文描述了一种由微型计算机控制的泵,该泵能产生类似于人体外周循环中的脉动血流。使用连续波多普勒超声来研究该泵产生的血流,并检查从频谱得出的诊断指标的行为。从具有不同程度轴对称狭窄的弹性管和刚性管记录了声谱图。依次改变心率、心搏量、血管狭窄程度、血管壁弹性、远端阻力和全身外周阻抗,并比较由此产生的多普勒频谱。所考虑的指标有搏动指数、频谱展宽指数和峰值多普勒频率。总体而言,这些指标缺乏检测低至中度疾病水平的敏感性。