Using the Real-Ear-to-Coupler Difference within the American Academy of Audiology Pediatric Amplification Guideline: Protocols for Applying and Predicting Earmold RECDs.

作者信息

Moodie Sheila, Pietrobon Jonathan, Rall Eileen, Lindley George, Eiten Leisha, Gordey Dave, Davidson Lisa, Moodie K Shane, Bagatto Marlene, Haluschak Meredith Magathan, Folkeard Paula, Scollie Susan

出版信息

J Am Acad Audiol. 2016 Mar;27(3):264-275. doi: 10.3766/jaaa.15086.

Abstract

BACKGROUND

Real-ear-to-coupler difference (RECD) measurements are used for the purposes of estimating degree and configuration of hearing loss (in dB SPL ear canal) and predicting hearing aid output from coupler-based measures. Accurate measurements of hearing threshold, derivation of hearing aid fitting targets, and predictions of hearing aid output in the ear canal assume consistent matching of RECD coupling procedure (i.e., foam tip or earmold) with that used during assessment and in verification of the hearing aid fitting. When there is a mismatch between these coupling procedures, errors are introduced.

PURPOSE

The goal of this study was to quantify the systematic difference in measured RECD values obtained when using a foam tip versus an earmold with various tube lengths. Assuming that systematic errors exist, the second goal was to investigate the use of a foam tip to earmold correction for the purposes of improving fitting accuracy when mismatched RECD coupling conditions occur (e.g., foam tip at assessment, earmold at verification).

STUDY SAMPLE

Eighteen adults and 17 children (age range: 3-127 mo) participated in this study.

DATA COLLECTION AND ANALYSIS

Data were obtained using simulated ears of various volumes and earmold tubing lengths and from patients using their own earmolds. Derived RECD values based on simulated ear measurements were compared with RECD values obtained for adult and pediatric ears for foam tip and earmold coupling.

RESULTS

Results indicate that differences between foam tip and earmold RECDs are consistent across test ears for adults and children which support the development of a correction between foam tip and earmold couplings for RECDs that can be applied across individuals.

CONCLUSIONS

The foam tip to earmold correction values developed in this study can be used to provide improved estimations of earmold RECDs. This may support better accuracy in acoustic transforms related to transforming thresholds and/or hearing aid coupler responses to ear canal sound pressure level for the purposes of fitting behind-the-ear hearing aids.

摘要

背景

真耳到耦合器差值(RECD)测量用于估计听力损失的程度和形态(以dB SPL耳道为单位),并根据基于耦合器的测量结果预测助听器输出。准确测量听力阈值、推导助听器适配目标以及预测耳道内的助听器输出,均假定RECD耦合程序(即泡沫耳塞或耳模)与评估及助听器适配验证期间所使用的程序一致匹配。当这些耦合程序不匹配时,就会引入误差。

目的

本研究的目的是量化使用泡沫耳塞与不同管径耳模时测得的RECD值的系统差异。假设存在系统误差,第二个目的是研究在RECD耦合条件不匹配(例如,评估时使用泡沫耳塞,验证时使用耳模)的情况下,使用从泡沫耳塞到耳模的校正来提高适配精度。

研究样本

18名成年人和17名儿童(年龄范围:3 - 127个月)参与了本研究。

数据收集与分析

使用不同体积的模拟耳和不同管径的耳模管,并从使用自己耳模的患者那里获取数据。将基于模拟耳测量得出的RECD值与成人和儿童耳朵使用泡沫耳塞和耳模耦合时获得的RECD值进行比较。

结果

结果表明,成人和儿童受试耳的泡沫耳塞和耳模RECD之间的差异是一致的,这支持了开发一种可应用于个体的泡沫耳塞和耳模耦合之间的RECD校正方法。

结论

本研究中开发的从泡沫耳塞到耳模的校正值可用于更准确地估计耳模RECD。这可能有助于提高与转换阈值和/或助听器耦合器响应到耳道声压级相关的声学转换的准确性,以便适配耳背式助听器。