Baart A Mireille, de Kort Wim L A M, van den Hurk Katja, Pasker-de Jong Pieternel C M
Department of Donor Studies, Sanquin Research, Amsterdam, the Netherlands.
Transfusion. 2016 Aug;56(8):1984-93. doi: 10.1111/trf.13546. Epub 2016 Mar 9.
Most donors do not like the invasive hemoglobin (Hb) measurement in fingerstick capillary samples. Alternative noninvasive devices have recently become available. In this study we assessed the precision and practicability of noninvasive devices and compared them with measurements in capillary and venous samples.
A total of 2589 blood donors were included in the study. Hb levels were measured with the noninvasive devices NBM 200 or Haemospect and compared with values obtained with the invasive devices HemoCue 201 and HemoCue 301 and with venous values. Precision was evaluated in terms of sensitivity and specificity for detection of Hb levels below the cutoff level for donation, taking venous values as "reference values."
Hb levels measured with both invasive and noninvasive devices differed significantly from venous measurements. Measurements with invasive devices correlated stronger with venous values and were more precise. Sensitivity ranged from 30.1% (HemoCue 201) to 3.5% (Haemospect) in men and from 36.6% (HemoCue 201) to 10.8% (Haemospect) in women. Specificity ranged from 99.8% (HemoCue 201) to 95.5% (Haemospect) in men and from 98.9% (HemoCue 201) to 94.6% (NBM 200) in women. Most donors preferred the noninvasive measurement.
Measurements with invasive devices showed stronger correlation and better agreement with venous values than noninvasive devices. Sensitivity was low for all measurement devices, particularly the noninvasive ones. In terms of precision invasive measurement would be preferred for donor screening, although a majority of donors preferred the noninvasive measurement.
大多数献血者不喜欢通过采集指尖毛细血管样本进行侵入性血红蛋白(Hb)测量。近年来出现了替代的非侵入性设备。在本研究中,我们评估了非侵入性设备的精度和实用性,并将其与毛细血管和静脉样本的测量结果进行比较。
共有2589名献血者纳入本研究。使用非侵入性设备NBM 200或Haemospect测量Hb水平,并与使用侵入性设备HemoCue 201和HemoCue 301获得的值以及静脉血值进行比较。以静脉血值作为“参考值”,通过检测低于献血截止水平的Hb水平的敏感性和特异性来评估精度。
侵入性和非侵入性设备测量的Hb水平与静脉血测量值均存在显著差异。侵入性设备的测量结果与静脉血值的相关性更强且更精确。男性的敏感性范围为30.1%(HemoCue 201)至3.5%(Haemospect),女性为36.6%(HemoCue 201)至10.8%(Haemospect)。男性的特异性范围为99.8%(HemoCue 201)至95.5%(Haemospect),女性为98.9%(HemoCue 201)至94.6%(NBM 200)。大多数献血者更喜欢非侵入性测量。
与非侵入性设备相比,侵入性设备的测量结果与静脉血值的相关性更强且一致性更好。所有测量设备的敏感性都较低,尤其是非侵入性设备。在精度方面,尽管大多数献血者更喜欢非侵入性测量,但侵入性测量更适合用于献血者筛查。