Date-Chong Mavis, Hudlow William R, Buoncristiani Martin R
Jan Bashinski DNA Laboratory, Bureau of Forensic Services, California Department of Justice, Richmond, CA 94804, USA.
Forensic Sci Int Genet. 2016 Jul;23:1-8. doi: 10.1016/j.fsigen.2016.03.001. Epub 2016 Mar 3.
The RapidHIT™ 200 Human Identification System and RapidHIT GlobalFiler(®) Express kit were evaluated and validated for use with single-source reference samples. It was of primary interest to evaluate the system for its efficacy as an expert system and to estimate a first pass success rate, as well as to identify the technical variables impacting that result. While results indicated that this instrument/kit combination can be used to accurately type single-source buccal samples, substantial variability in sensitivity and intra-color balance were observed, as were multiple artifacts, requiring extensive manual editing of the profiles. Artifacts included dye "blobs" and spectral overlap (pull-up) peaks that often originated from relatively low intensity allele peaks. Reduced intra-color balance, in combination with low sensitivity, occasionally resulted in instances of allelic dropout. Overall, 50% of the buccal samples analyzed in this study would have been successfully typed to give full GlobalFiler(®) profiles without the need for manual review and editing.
对RapidHIT™ 200人类识别系统和RapidHIT GlobalFiler(®) Express试剂盒进行了评估和验证,以用于单源参考样本。主要目的是评估该系统作为专家系统的有效性,估计首次通过率,并确定影响该结果的技术变量。虽然结果表明该仪器/试剂盒组合可用于准确分型单源颊拭子样本,但观察到灵敏度和颜色内平衡存在很大差异,还存在多个伪像,需要对图谱进行大量手动编辑。伪像包括染料“斑点”和光谱重叠(拖尾)峰,这些峰通常源自强度相对较低的等位基因峰。颜色内平衡降低,再加上灵敏度较低,偶尔会导致等位基因缺失的情况。总体而言,本研究中分析的50%的颊拭子样本无需人工审核和编辑就能成功分型,得出完整的GlobalFiler(®)图谱。