体重≤75kg患者腹部CT检查期间将管电压从120kVp手动调整至100kVp:一项前瞻性随机试验中的图像质量与辐射剂量评估

Manual adjustment of tube voltage from 120 to 100 kVp during abdominal CT in patients with body weights ≤75 kg: assessment of image quality and radiation dose in a prospective, randomised trial.

作者信息

Zaehringer C, Euler A, Karwacki G M, Hohmann J, Pansini M, Szucs-Farkas Z, Schindera S T

机构信息

Clinic of Radiology and Nuclear Medicine, University Hospital Basel, Basel, Switzerland.

Department of Diagnostic Radiology, Hospital Center of Biel, Biel, Switzerland.

出版信息

Clin Radiol. 2016 Jun;71(6):615.e1-6. doi: 10.1016/j.crad.2016.02.014. Epub 2016 Mar 19.

Abstract

AIM

To assess image quality and radiation dose in patients with body weights ≤75 kg undergoing abdominal computed tomography (CT) with a tube voltage of either 120 or 100 kVp.

MATERIALS AND METHODS

Eighty patients weighing ≤75 kg were prospectively assigned to receive either 120 or 100 kVp abdominal CT in the portal-venous phase. Attenuation values of abdominal organs and image noise were measured, and the contrast-to-noise ratios (CNRs) were calculated. Subjective image quality was assessed by three independent radiologists. Radiation exposure was assessed by size-specific dose estimate (SSDE).

RESULTS

The mean attenuation of the kidney increased by 20% at 100 kVp (p<0.0001), and the mean image noise was 27% higher in the 100 kVp (p=0.003). The CNR did not significantly differ between the groups (120 kVp, 6.6±2.8; 100 kVp, 7.4±3.6; p=0.26). Except for subjective image noise (p<0.001), no other subjective quality parameters (e.g., contrast, artefacts) were significantly different between the two groups (p between 0.094 and 0.761). The mean SSDE in the 100-kVp group (9.8±1.8 mGy) was reduced by 19% compared to the 120-kVp group (12.1±1.8 mGy; p<0.0001).

CONCLUSION

Manual reduction of tube voltage from the standard 120 to 100 kVp for portal-venous phase CT in patients with body weights ≤75 kg resulted in a 19% dose reduction while maintaining objective and subjective image quality.

摘要

目的

评估体重≤75kg的患者在进行管电压为120kVp或100kVp的腹部计算机断层扫描(CT)时的图像质量和辐射剂量。

材料与方法

前瞻性地将80名体重≤75kg的患者分配接受门静脉期120kVp或100kVp的腹部CT检查。测量腹部器官的衰减值和图像噪声,并计算对比噪声比(CNR)。由三名独立的放射科医生评估主观图像质量。通过特定尺寸剂量估计(SSDE)评估辐射暴露。

结果

100kVp时肾脏的平均衰减增加了20%(p<0.0001),100kVp时的平均图像噪声高27%(p=0.003)。两组之间的CNR没有显著差异(120kVp,6.6±2.8;100kVp,7.4±3.6;p=0.26)。除主观图像噪声(p<0.001)外,两组之间没有其他主观质量参数(如对比度、伪影)有显著差异(p在0.094至0.761之间)。100kVp组的平均SSDE(9.8±1.8mGy)比120kVp组(12.1±1.8mGy)降低了19%(p<0.0001)。

结论

对于体重≤75kg的患者,在门静脉期CT检查时将管电压从标准的120kVp手动降低到100kVp可使剂量降低19%,同时保持客观和主观图像质量。